GE LOGIQ 500 Advanced Reference Manual page 241

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

S220
Mode
B-Mode
M-Mode
B/M-Mode
Pulsed Wave Doppler
B/Pulsed Wave Doppler
B/Color Flow Doppler
B/Color Flow Doppler/
Pulsed Wave Doppler
M-Mode/Color M-Mode
Color Flow with M Doppler
CW Doppler
Table 6–48. S220 Maximum Acoustic Output Levels
S222
Mode
B-Mode
M-Mode
B/M-Mode
Pulsed Wave Doppler
B/Pulsed Wave Doppler
B/Color Flow Doppler
B/Color Flow Doppler/
Pulsed Wave Doppler
M-Mode/Color M-Mode
Color Flow with M Doppler
CW Doppler
Table 6–49. S222 Maximum Acoustic Output Levels
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
TIS/TIB
MI
Scan
0.51
0.44
0.51
0.51
0.29
0.58
0.58
0.29
0.83
1.12
0.83
0.71
0.83
0.83
0.03
TIS/TIB
TIS
MI
Scan
Non-Scan
1.44
0.85
1.44
0.17
1.44
0.57
0.17
0.70
0.22
1.44
< 0.1
0.22
1.44
2.05
1.44
1.03
0.11
1.44
0.21
1.44
0.22
0.15
2.20
FDA Acoustic Output Data
TIS
TIB
Non-Scan
Non-Scan
0.07
0.20
0.07
0.20
0.59
2.52
0.59
2.52
0.29
1.26
0.53
2.71
0.53
2.71
0.84
1.16
TIB
Non-Scan
TIC
1.97
1.75
0.65
1.75
1.96
2.27
0.84
2.27
0.87
2.41
1.14
1.63
4.01
0.86
2.27
0.84
2.25
4.32
TIC
I
SPTA.3
0.64
201.50
0.10
53.80
0.52
188.13
1.12
334.94
1.55
469.27
1.35
462.63
1.45
465.95
0.93
496.88
0.93
496.88
2.17
183.66
I
I
SPTA.3
SPPA.3
382.4
587.8
446.8
587.8
701.7
587.8
674.2
141.6
680.0
587.8
346.0
587.8
512.0
587.8
578.2
587.8
674.2
587.8
406.6