S611 Probe
Parameter
P_ (MPa)
2
I
(mW/cm
)
spta
I
(mm)
p
w
pb6
(||) (mm)
( )(mm)
prr (kHz)
srr (Hz)
Output Beam Dim.
(||) (mm)
( )(mm)
f
(MHz)
awf
APF (%)
(Acoustic Power
up Fraction)
AIF (%)
(Acoustic
Initialization
Fraction)
Maximum Power (mW)
2
I
(mW/cm
)
ob
Power-up Mode
Initialization Mode
Acoustic Output
Freeze
I
(mm)
tt
I
(mm)
ts
Table 6–169. S611 IEC Acoustic Output Information
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
B
M
2.59
2.59
295.29
199.40
37.10
37.10
1.14
1.14
2.22
2.22
0.85
142.00
11.50
11.50
8.00
8.00
4.43
4.43
na
na
na
na
20.88
2.25
22.69
2.50
na
na
na
na
yes
yes
0.00
0.00
contact
contact
IEC Acoustic Output Tables
B+M
D
2.59
2.96
396.26
2358.70
37.10
45.00
1.14
1.33
2.22
1.94
0.85
1.49
142.00
11.50
11.20
8.00
8.00
4.43
4.98
na
na
na
na
16.17
40.55
17.63
44.56
na
na
na
na
yes
yes
0.00
0.00
contact
contact
B+D
2.96
2555.56
45.00
1.33
1.94
1.49
142.00
11.50
8.00
4.98
na
na
54.47
59.68
na
na
yes
0.00
contact
6–143