S317 Probe
Parameter
P_ (MPa)
2
I
(mW/cm
)
spta
I
(mm)
p
w
pb6
(||) (mm)
( )(mm)
prr (kHz)
srr (Hz)
Output Beam Dim.
(||) (mm)
( )(mm)
f
(MHz)
awf
APF (%)
(Acoustic Power
up Fraction)
AIF (%)
(Acoustic
Initialization
Fraction)
Maximum Power (mW)
2
I
(mW/cm
)
ob
Power-up Mode
Initialization Mode
Acoustic Output
Freeze
I
(mm)
tt
I
(mm)
ts
Table 6–168. S317 IEC Acoustic Output Information
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
B
M
2.70
2.70
821.8
892.5
46.7
46.7
2.0
2.0
3.0
3.0
0.86
82
12.0
12.0
12.9
12.9
2.63
2.63
na
na
na
na
148.24
24.57
95.91
15.90
na
na
na
na
yes
yes
0.0
0.0
contact
contact
IEC Acoustic Output Tables
B+M
D
2.70
1.96
1440.4
1495.8
46.7
66.0
2.0
2.3
3.0
3.9
0.86
1.71
82
12.0
12.0
12.9
17.9
2.63
2.51
na
na
na
na
123.39
70.54
79.84
32.81
na
na
na
na
yes
yes
0.0
0.0
contact
contact
B+D
2.70
2043.6
66.0
2.3
3.9
1.71
82
12.0
17.9
2.51
na
na
169.37
96.74
na
na
yes
0.0
contact
6–141