Lv Calculation Formulas (Teichholz Method) - GE LOGIQ 500 Advanced Reference Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

LV Calculation Formulas (Teichholz Method)

Figure 4–56. Teichholz Method Measurements
Calc Mnemonic
SI
Stroke Volume Index
CI
Cardiac Index
MVCF
Mean Vcf
LVM
Left Ventricle Cardiac
Mass
Table 4–3. Teichholz Method Calculation Formulas
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
B-Mode
Refer to Chapter 10 (Cardiology) of the Basic User Manual for
the following Teichholz Method Formulas:
LVIDd, LVIDs, IVSd, IVSs, LVPWd, LVPWs, HR, ET, EdV, EsV,
FS, SV, EF and CO.
Calc Name
Input Measurements
body surface area and SV
body surface area and CO
two distances & one time
interval
three distances
Advanced Cardiac Calculations
M-Mode
SI[ml/m^2]=SV/BSA
CI[ml/min/m^2]=
CO/BSA
MVCF[circ/s]=
(LVIDd-LVIDs)/
(LVIDd x ET)
LVM[g]=1.04*[(IVSd+
LVPWd+LVIDd)^3–
LVIDd^3]–13.6
Formula