Troubleshooting

Inspection Sequence
Operation Guide/Error Messages
No.
Message
(1)
Cannot read Tape ID.
Register the new tape?
(y/n)
(2)
Check VCR.
8–44
If trouble is suspected, conduct a check in the following order.
1. An operation guide/error message appears at the
bottom of the system monitor.
Refer to Operation Guide/Error Messages on 8–44 .
2. Check the items in A List of Problems and How To
Trouble Shoot Them on 8–47 .
3. Check the VCR switch settings.
Refer to Setting Up the VCR on 8–11 .
If the problem remains after conducting the above,
contact a qualified service representative.
The table below shows each message and the appropriate
action to take. The message column lists the operation
guide/error messages displayed at the bottom of the monitor
during VCR operation.
Possible Cause
1)
An unregistered new tape
was inserted into the VCR.
2)
The tape ID could not be
read.
There is no signal input or re-
sponse from the VCR.
Table 8–2. Operation Guide/Error Messages
Possible Corrective Action
Type "y" to register a new tape. Re-
fer to Registering a New Tape on
8–25 .
Type "n" to proceed to message (08)
to re-attempt reading the tape drive.
1)
Check the VCR power cord and
all the cable connections.
2)
Check the settings of the LO-
GIQ
500 System Parameters
menu page 5.
Refer to Customizing Your Sys-
tem in LOGIQ
Reference Manual.
3)
Check the VCR switch settings.
Refer to Setting Up the VCR on
8–11 .
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0