GE LOGIQ 500 Advanced Reference Manual page 53

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

Transferring Calcs
.
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
Selecting Transf. CALCs copies data from the Realtime
Doppler Calculation window to the selected location on the
report page.
The following is an example of transferring calculations in the
OB exam category:
1. Select OB PI, OB RI and OB S/D (D/S) from the
Doppler Realtime Calculation : Submenu selection in
the Set Up/Preset Program menu page 7.
2. Turn on the Realtime Doppler Calculation selection in
the Softmenu.
3. The selected measurements will be displayed in the
upper left portion of the screen.
4. Press Freeze.
5. Choose "Select Location" from the Measurement
Softmenu.
6. Select "Transf. CALCS" to transfer the calculations to
the selected location on the report page.
The Transf. CALCs function is not available unless values are
displayed in the Doppler Realtime Display window.
NOTE: The Advanced Vascular Calculation Sub-Menu has an
option called Manual Calc. When Manual Calc is activated,
Auto Calc measurements will not be transferred to the Report
Page or result window.
Realtime Doppler Calculations
4–7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents