Urology Calculations (Option)
ERB7 Probe Preparation (cont'd)
.
1
2
3
4
5
Figure 4–93. Sterile/Sanitary Sheath Placement
4–118
1. Remove the probe from the box and carefully examine
it for any damage.
2. Clean and then disinfect/sterilize the ERB7 Probe.
3. Inspect a sterile/sanitary sheath. Place a small amount
of ultrasound gel inside the sheath tip (the gel is
between the sheath inner surface and the probe
aperture).
4. Place the sheath tip over the probe aperture and then
pull the sheath end toward the probe handle.
5. Place a rubberband/twist lock over the sheath at the
end of the probe shaft. Ensure the rubberband/twist
lock is tight around the sheath. Rub your finger over
the tip of the probe to ensure all air bubbles have been
eliminated.
6. Place a small amount of ultrasound gel on the gel-filled
sheath tip outer surface.
NOTE: Remember to rinse all probe sheaths of powder before
placing on the probe. Powder can degrade the scan
image displayed.
1
Twist Lock
Sterile/Sanitary Outer Sheath
Optional Finger Cot for Water Path
Optional Water Path Inlet
Optional Finger Cot Rubberband for Water Path
2
4
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
3
5