Operation Guide/Error Messages (cont'd)
No.
Message
(3)
Check VCR. No Cas-
sette.
(4)
Please play VCR
image and freeze at
the other frame.
(5)
Protected. Check
Tape.
(6)
Requested information
doesn't exist on the
tape.
(7)
Requested information
doesn't exist in Sys-
tem.
(8)
Searching Tape ID and
System ID
(9)
Not registered on this
system. Can't record.
(10)
Rewinding to begining
for Patient Info.
(11)
Reading Patient In-
formation
(12)
Returning to tape inser-
tion position
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
Possible Cause
The VCR has no tape.
The image information on a tape
cannot be read.
This tape has been write pro-
tected.
The patient image is not on the
tape.
The patient information is not on
the hard drive.
After Tape insertion system
searches
Tape ID and system ID do not
match
Displayed during rewind after #9
is displayed
System reading info from begin-
ing of tape
Tape not recognized so system
fast forwards to tape position
when it was inserted
Table 8–2. Operation Guide/Error Messages (cont'd)
VCR Operating Instructions
Possible Corrective Action
Insert a tape into the VCR.
Advance the tape a few frames, then
freeze the image.
Use an un-protected tape.
1)
Verify search conditions and re-
attempt, if necessary.
2)
When an inappropriate tape is
used, cancel image search and
press Ctrl and T to activate a
tape search. Find the tape on
which the relevant patient de-
tected by the tape search has
been recorded. Insert the rele-
vant tape into the VCR and re-
attempt an image search. Refer
to Advanced Search on 8–34 .
1)
Verify search conditions and re-
attempt, if necessary.
2)
When an inappropriate system is
used, cancel tape search and re-
attempt the search on the correct
system. Refer to Tape Search on
8–38 .
8–45