GE LOGIQ 500 Advanced Reference Manual page 79

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

Error Messages (cont'd)
Hints
.
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
S
Data transfer can only be accomplished in the OB, GYN
and Vascular Exam categories.
S
Date is always transferred as Year–Month–Day. Time is
transferred as 24 hour time (i.e. 00:00 to 23:59).
S
Gestational Age Data is always transferred in weeks and
days.
S
The LOGIQ
500 can store three values of the same
measurement. Only the average of selected
measurements will be transferred.
S
The system keyboard is locked out during the transfer of
data.
S
Only the Record 1 or Record 2 keys can start the data
transfer.
S
The following DMC items are not sent by the LOGIQ
S
Serial No.
S
DateRep
S
TimeRep
S
RepPtName
S
RefMDID
S
History
S
FetusPos
S
FetusSex.
NOTE: Currently, the transfer of multi-fetus information is not
compatible with the DMC. If it is desired to use the DMC with
multiple fetuses, each fetus should be treated individually by
using a unique Patient ID number for each.
Data Management Center (DMC)