Setting up a Record Keeping System

Preparation
Record Keeping
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
The following is needed:
S
Quality Assurance binder
S
Magnetic photo pages for hard copy images, or Floppy disk
holder for archived images
S
Quality Assurance Checklists.
Display the following information while testing quality
assurance:
S
Acoustic Output
S
Gain
S
Depth
S
Probe
S
Dynamic Range
S
Set up new patient to be the name of the test.
Annotate the following:
S
Any control where its value is NOT displayed.
S
Significant phantom information.
Complete the following:
1. Fill out the Ultrasound Quality Assurance Checklist for
each probe, as scheduled.
2. Make a hard copy or archive the image.
3. Compare images to baseline images and acceptable
values.
4. Evaluate trends over previous test periods.
5. File hard copy/archive floppy and checklist in Quality
Assurance binder.
Quality Assurance