FDA Acoustic Output Data
LD
Mode
B-Mode
M-Mode
B/M-Mode
Pulsed Wave Doppler
B/Pulsed Wave Doppler
B/Color Flow Doppler
B/Color Flow Doppler/
Pulsed Wave Doppler
M-Mode/Color M-Mode
Color Flow with M Doppler
P509
Mode
B-Mode
M-Mode
B/M-Mode
Pulsed Wave Doppler
B/Pulsed Wave Doppler
B/Color Flow Doppler
B/Color Flow Doppler/
Pulsed Wave Doppler
M-Mode/Color M-Mode
Color Flow with M Doppler
CWD
Table 6–47. P509 Maximum Acoustic Output Levels
6–56
TIS/TIB
MI
Scan
0.63
0.21
0.63
0.63
0.14
0.58
0.63
0.14
0.90
1.35
0.90
0.74
0.79
0.79
Table 6–46. LD Maximum Acoustic Output
TIS/TIB
MI
Scan
0.88
0.25
0.88
0.88
0.16
0.81
0.88
0.16
0.88
0.63
0.88
0.40
0.88
0.88
< 0.1
TIS
TIB
Non-Scan
Non-Scan
<0.1
<0.1
<0.1
<0.1
0.33
2.32
0.33
2.32
0.16
1.16
0.29
2.13
0.29
2.13
TIS
TIB
Non-Scan
Non-Scan
< 0.1
0.18
< 0.1
0.18
0.56
1.63
0.56
1.63
0.28
0.82
0.48
1.56
0.48
1.56
1.33
1.73
LOGIQ
500 Advanced Reference Manual
TIC
I
SPTA.3
0.49
4.4
<0.1
26.1
0.38
29.1
1.70
680.6
2.02
683.6
1.37
82.6
1.69
383.1
1.23
655.5
1.23
655.5
TIC
I
SPTA.3
0.26
287.1
< 0.1
43.9
0.26
235.3
0.66
405.9
0.83
597.4
0.72
470.7
0.77
534.0
0.54
455.4
0.54
455.4
2.49
232.8
2276614–100 Rev. 0