Lv Calculation Formulas (Lv Bp-Disc Method) - GE LOGIQ 500 Advanced Reference Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

Advanced Cardiac Calculations

LV Calculation Formulas (LV BP-DISC Method)

LVLs
LVAs
Figure 4–59. LV BP-DISC Method Measurements
Calc Mnemonic
LVA2d
Left Ventricular Area,
Apical 2 Chamber, Diastole
LVA2s
Left Ventricular Area,
Apical 2 Chamber, Systole
LVA4d
Left Ventricular Area,
Apical 4 Chamber, Diastole
LVA4s
Left Ventricular Area,
Apical 4 Chamber, Systole
LVL2d
Left Ventricular Length,
Apical 2 Chamber, Diastole
LVL2s
Left Ventricular Length,
Apical 2 Chamber, Systole
LVL4d
Left Ventricular Length,
Apical 4 Chamber, Diastole
LVL4s
Left Ventricular Length,
Apical 4 Chamber, Systole
HR
Heart Rate
(beats/minute)
4–70
LVLs
LVAs
Calc Name
Input Measurements
one area (by trace only) is
divided into 20 elliptical
disk segments
one area (by trace only) is
divided into 20 elliptical
disk segments
one area (by trace only) is
divided into 20 elliptical
disk segments
one area (by trace only) is
divided into 20 elliptical
disk segments
one distance
one distance
one distance
one distance
one 2 beat time interval
Table 4–6. LV BP-DISC Method Formulas
LVLd
LVAd
LVA2d=a1[cm ^2]
LVA2s=a1[cm ^2]
LVA4d=a1[cm ^2]
LVA4s=a1[cm ^2]
LVL2d=d1[cm or mm]
LVL2s=d1[cm or mm]
LVL4d=d1[cm or mm]
LVL4s=d1[cm or mm]
HR[BPM]=120 [sec]/ 2
beat time [sec]
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
LVLd
LVAd
Formula

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents