Servicing; Intervals And Scope - Weinmann MEDUMAT Basic Servicing And Repair Instructions

Ventilators
Table of Contents

Advertisement

4. Servicing

N.B.
Always remember to carry out a technical safety check of the ventilator after every repair.
MEDUMAT Basic / Basic p must be serviced regularly.
We recommend having all maintenance work, servicing and repairs carried out either by the manufacturer
Weinmann or by a qualified agent expressly authorised by that company.

4.1 Intervals and Scope

Every 2 years:
Every 2 years, you must subject the device (including patient valve and tube system) to a
and maintenance.
The servicing and inspection may also be carried out by the manufacturer Weinmann.
The following points should be observed:
Check that the equipment is complete
Visual check for:
– physical or mechanical damage
– correct markings on controls
– damage to all external hoses
Replacement of worn components/
compulsory change parts (see "7.2 Mainte-
nance set" on page 49);
Check of system components: portable sys-
tem, oxygen supply fittings, secretion suction
system, hose connections etc.
Every 4 years:
Servicing of the fittings in the oxygen supply system (e.g. pressure reducer) either by the manufacturer or
by a qualified agent expressly authorised by him.
Every 10 years:
Repeat testing of the conventional steel or aluminium oxygen bottles by the Technical Testing Association.
The specified testing date is stamped on the shoulder of the bottle.
Check test bag.
Repeat testing of aluminium oxygen bottles
WM 1821 and WM 3621 by the Technical
Testing Association. The specified testing date
is stamped on the shoulder of the bottle.
Final check in accordance with Test Instructions/
Test Report STK WM 22671 (see „3. Final Check"
on page 8 and see "11. Repair and inspection
log" on page 57).