TELEDYNE Model 9110AH NO

5.2 Sample Mode

5.2.1 Test Functions

In any of the following TEST functions, if a value of
TEST functions allow the operator to examine several important operating parameters of the
M9110AH to verify correct Analyzer operation. If faults are discovered, refer to
Troubleshooting Section 9.
Range (RANGE)
This is the Range of the instrument. In single range mode there is one range for all 3 rear panel
analog outputs.
The AutoRange option allows 2 different ranges for each channel, and will automatically switch
to the other range dynamically as concentration values require. By selecting Remote
AutoRange, the range change can be controlled remotely. The TEST values will show the range
the instrument is currently operating in, and will dynamically display range change information
when necessary.
Independent range option allows different ranges for each output. When enabled, there will be
three range values displayed, NO, NO
Each of the range modes Single range, Auto range, and
Signal Stability (STABIL)
The instrument noise is computed by calculating the standard deviation of the last 10 minutes of
concentration data. If NO/ NO
channel. If NO only or NO
noise value only becomes meaningful if sampling a constant concentration for more than 10
minutes. The noise value should be compared to the value observed in the factory check-out.
Analyzer Operator Manual, 01620, Rev. F
X
NOTE
XXXX is displayed, that indicates an off scale and
therefore meaningless reading.
x
NOTE
Independent ranges are mutually exclusive.
switching mode is selected, the noise reading is that of the NO
x
only mode is selected the Noise applies to the mode selected. The
x
and NO
.
2
5-9