Ain Calibration; Table 6-12: Current Loop Output Calibration With Resistor - Teledyne 100E Instruction Manual

Uv fluorescence so2 analyzer
Table of Contents

Advertisement

Operating Instructions
If a current meter is not available, an alternative method for calibrating the current loop
outputs is to connect a 250 Ω ±1% resistor across the current loop output. Using a
voltmeter, connected across the resistor, follow the procedure above but adjust the
output to the following values:

Table 6-12: Current Loop Output Calibration with Resistor

Full scale
0%
100%

6.7.4.6. AIN Calibration

This is the sub-menu to conduct the analog input calibration. This calibration should only
be necessary after major repair such as a replacement of CPU, motherboard or power
supplies. Activate the ANALOG I/O CONFIGURATION MENU (see Section 6.7.1), then
press:
Instrument
calibrates
automatically
84
Voltage for 2-20 mA
(measured across resistor)
0.5 V
5.0 V
STARTING FROM ANALOG I / O CONFIGURATION MENU
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
Continue pressing SET? until ...
DIAG AIO
AIN CALIBRATED:
< SET SET>
CAL
DIAG AIO
CALIBRATING A/D ZERO
CALIBRATING A/D SPAN
DIAG AIO
AIN CALIBRATED:
< SET SET>
CAL
Model 100E Instruction Manual
Voltage for 4-20 mA
(measured across resistor)
1.0 V
5.0 V
EXIT
ENTR
NO
EXIT
YES
EXIT
Exit at any time to
return to the main
DIAG menu
Exit to return to the
ANALOG I/O
CONFIGURATION
MENU
045150102 Rev XB1