Operating Instructions

Table 6-12: Current Loop Output Calibration with Resistor

Full scale
0%
100%

6.7.3.6. AIN Calibration

This is the sub-menu to conduct the analog input calibration. This calibration should only be
necessary after major repair such as a replacement of CPU, motherboard or power supplies.
Instrument
calibrates
automatically

6.7.4. Test Channel Output

When activated, output channel A4 can be used to report one of the test functions viewable
from the SAMPLE mode display.
68
Voltage for 2-20 mA
(measured across resistor)
0.5 V
5.0 V
STARTING FROM ANALOG I / O CONFIGURATION MENU
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
DIAG AIO
AIN A/C FREQUENCY: 60 HZ
SET>
EDIT
DIAG AIO
AIN CALIBRATED:
< SET SET>
CAL
DIAG AIO
CALIBRATING A/D ZERO
CALIBRATING A/D SPAN
DIAG AIO
AIN CALIBRATED:
< SET SET>
CAL
Model 200E Instruction Manual
Voltage for 4-20 mA
(measured across resistor)
1.0 V
5.0 V
Exit at any time to
return to the main
DIAG menu
ENTR
EXIT
EXIT
NO
EXIT
Exit to return to the
YES
CONFIGURATION
EXIT
ANALOG I/O
MENU
044100102 Rev A