11.5.7.3. Analog Outputs: Current Loop
To verify that the Analog outputs with the optional current mode output are
working properly, connect a 250 ohm resistor across the outputs and use a
voltmeter to measure the output as described in Section 6.5.2 and then perform an
Analog Output Step Test as described in Section 6.6.2.
For each step the output should be within 1% of the nominal value listed in the
table below.

Table 11-9: Analog Output Test Function - Nominal Values Current Outputs

Step
1
2
3
4
5
6
11.5.7.4. Status Outputs
The procedure below can be used to test the Status outputs:
18. Connect a jumper between the "D" pin and the "s" pin on the status output
connector.
19. Connect a 1000 ohm resistor between the "+" pin and the pin for the status
output that is being tested.
20. Connect a voltmeter between the "s" pin and the pin of the output being
tested (see table below).
21. Under the DIAG:Signal I/O menu (see Section 11.1.3), scroll through the
inputs and outputs until you get to the output in question. Alternately turn on
and off the output noting the voltage on the Voltmeter, it should vary between
0 volts for ON and 5 volts for OFF.
P/N 04288 REV: A
%
Current
0
2 mA
20
5.6
40
9.2
60
12.8
80
16.4
100
20
TROUBLESHOOTING & REPAIR PROCEDURES
Output Range
2 -20
Nominal Output Values
V(250 Ohms)
0.5V
1.4
2.3
3.2
4.1
5
4 -20
Current
V(250 Ohms)
4
1
7.2
1.8
10.4
2.6
13.6
3.4
16.8
4.2
20
5
213