Figure 6-3: Setup For Calibrating Current Outputs - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Model 200E Instruction Manual

Figure 6-3: Setup for Calibrating Current Outputs

Do not exceed 60 V between current loop outputs and instrument ground.
To adjust the zero and span values of the current outputs, press SETUP-MORE-DIAG-ENTR-
NEXT-NEXT and then:
FROM ANALOG I/O CONFIGURATION MENU
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
DIAG AIO
AIN A/C FREQUENCY: 60 HZ
SET>
EDIT
DIAG AIO
AIN CALIBRATED: NO
SET>
EDIT
DIAG AIO
AOUT CALIBRATED: NO
< SET
SET>
CAL
Press SET> to select the analog output channel
to be configured:. Then press EDIT to continue
DIAG AIO
CONC_OUT_CURR, NO CAL
< SET SET>
EDIT
DIAG AIO
CONC_OUT_2 RANGE: CURR
<SET
SET>
EDIT
If a current meter is not available, an alternative method for calibrating the current loop
outputs is to connect a 250 Ω ±1% resistor across the current loop output. Using a
voltmeter, connected across the resistor, follow the procedure above but adjust the output
to the following values:
044100102 Rev A
See Table 3-2 for
pin assignments of
the Analog Out
connector on the
rear panel.
V OUT +
V OUT -
Analyzer
CAUTION
DIAG AIO
< SET
ENTR
EXIT
DIAG AIO
U100
UP10
EXIT
DIAG AIO
U100 UP10 UP DOWN DN10 D100
EXIT
DIAG AIO
EXIT
U100 UP10 UP DOWN DN10
DIAG AIO
U100 UP10 UP DOWN DN10 D100
EXIT
DIAG AIO
< SET
EXIT
mA
IN
OUT
I IN +
I IN -
Recording
Device
CONC_OUT_2 CALIBRATED: NO
CAL
EXIT
CONC_OUT_2 ZERO: 0 mV
UP DOWN DN10 D100 ENTR EXIT
EXAMPLE
CONC_OUT_2 ZERO: 27 mV
ENTR EXIT
CONC_OUT_2 SPAN: 10000 mV
D100
ENTR EXIT
EXAMPLE
CONC_OUT_2 ZERO: 9731 mV
ENTR EXIT
CONC_OUT_2 CALIBRATED: YES
CAL
EXIT
Operating Instructions
Increase or decrease the current
output by 100, 10 or 1 counts. The
resulting change in output voltage is
displayed in the upper line.
Continue adjustments until the correct
current is measured with the current
meter.
ENTR returns
to the previous
menu.
EXIT ignores the
new setting, ENTR
accepts the new
setting.