Setting Other Trigger Parameters

These are trigger system parameters that apply in all trigger modes.

To set the trigger sweep

Trigger sweep specifies whether acquisitions occur without a triggern or
only with a trigger.
1 Press [Menu].
2 In the Trigger menu, press Sweep.
3 Continue pressing the Sweep softkey or turn the

To set the trigger coupling

Trigger coupling is used to filter low frequency signal components or DC
offsets from the trigger path when they interfere with achieving stable
triggers.
Trigger coupling is similar to channel coupling (see
affects the triggering system and does not change how the signal is
displayed.
To set the trigger coupling:
1 Press [Menu].
2 In the Trigger menu, press Set Up.
3 In the Set Up menu, press Coupling.
Agilent 1000B Series Oscilloscopes User's Guide
one of these trigger sweep settings:
Auto
Acquire waveform even when no trigger occurs.
Normal
Acquire waveform when trigger occurs.
Capturing Data
entry knob to select
page
47), but it only
3
93