Cleaning The Source Components - Waters Xevo TQD IVD Overview And Maintenance Manual

Table of Contents

Advertisement

Result:
The turbomolecular pump cover plate is now accessible.
9.
If the fan panel is obstructing access to the turbomolecular pump cover plate, disconnect
the fans power cable from the fan panel.
10. Follow the instructions in the Pfeiffer manual (Pfeiffer TurboDrag
remove the turbomolecular pump cover plate, replace the oil reservoir and porex rods, and
reattach the cover plate.
Tip:
The oil reservoir is referred to as the "operating fluid reservoir" in Pilfer's
documentation.
11. If you disconnected the fans power cable in
12. Carefully refit the fan panel to the instrument chassis.
13. Use the 4 mm screwdriver to refit the screw that secures the fan panel to the instrument
chassis.
Notice:
fitted correctly. The cover affects the flow of air over the turbomolecular pump.
14. Slide the lower front cover on to the instrument, then use the 3 mm screwdriver to refit the 7
screws that secure the cover.
15. If you removed the reservoir bottles in
16. Reconnect the power cable to the instrument's rear panel.
17. Refit the source enclosure onto the instrument (see
18. Start the mass spectrometer

4.17 Cleaning the source components

Clean the sample cone and cone gas nozzle when these conditions apply:
The sample cone and cone gas nozzle are visibly fouled.
Turbomolecular pump cover plate
To prevent the turbomolecular pump overheating, ensure that the cover is
step
(page
31).
September 7, 2016, 715003560IVD Rev. C
Page 63
®
Pump, HiPace 300) to
step
9, reconnect the cable to the fan panel.
4, refit the bottles to the fluidics system.
page
46).