Cleaning The Trizaic Source And Connectors - Waters TRIZAIC UPLC System Manual

Table of Contents

Advertisement

4.
Press the O-ring until it snaps into place around the base of the
compression fitting.
5.
Route the loose (unfitted) end of the fluid line through the fluid line
aperture and out of the source enclosure (see
page
2-8).
You can open and close the aperture by pulling and releasing the
Tip:
captive aperture assembly as indicated below.
6.
Fit a ferrule, locking ring, and compression fitting to the loose end of the
fluid line in accordance with the nanoACQUITY UPLC System
documentation.
See also:
7.
Attach the fluid line to the appropriate port on the nanoACQUITY
UPLC instrument, or to the onboard IntelliStart Fluidics on the mass
spectrometer (see

Cleaning the TRIZAIC source and connectors

Required materials
Chemical-resistant, powder-free gloves
Lint-free tissue
Water
Isopropanol or methanol
Source of clean, pressurized air
Caution:
surfaces.
During normal operation, the TRIZAIC source does not require cleaning. After
repeated use, however, material can accumulate around the fluid-line
connectors or electronic connectors in the nanoTile docking port.
nanoACQUITY UPLC System Operator's Guide.
"Starting up the system" on page
Do not use abrasives or solvents to clean the instrument's
Cleaning the TRIZAIC source and connectors
"Fluid line aperture:" on
2-9).
3-7

Advertisement

Table of Contents
loading

Table of Contents