Disassembling The Stepwave Ion Guide Assembly - Waters Xevo TQ-S Operator's, Overview And Maintenance Manual

Mass spectrometry system
Table of Contents

Advertisement

5.
Remove the StepWave ion guide assembly from the StepWave removal
and insertion tool.
6.
Using both hands, fit the source enclosure to the two supporting studs
on the source adaptor housing.
7.
Close the source enclosure.
Rationale:
entering the instrument while you are working on the StepWave ion
guide assembly.

Disassembling the StepWave ion guide assembly

Required materials
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
3-mm Allen wrench
Small, flat-blade screwdriver
Clean, lint-free cloth
4-64
Maintenance Procedures
The StepWave ion guide assembly is fragile; handle it
Caution:
with care.
Fitting and closing the source enclosure prevents debris