Disassembling The Stepwave Ion Guide Assembly - Waters Xevo TQ-S Operator's, Overview And Maintenance Manual

Ivd mass spectrometry system
Table of Contents

Advertisement

Caution:
with care.
5.
Remove the StepWave ion guide assembly from the StepWave removal
and insertion tool.
6.
Using both hands, fit the source enclosure to the two supporting studs
on the source adaptor housing.
7.
Close the source enclosure.
Rationale:
entering the instrument while you are working on the StepWave ion
guide assembly.

Disassembling the StepWave ion guide assembly

Required materials
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
3-mm Allen wrench
Small, flat-blade screwdriver
Clean, lint-free cloth
The StepWave ion guide assembly is fragile; handle it
Fitting and closing the source enclosure prevents debris
Cleaning the StepWave ion guide assembly
4-57