Removing The Measurement Bench And Lowering The Partition Panel - Thermo Scientific 49i Instruction Manual

Uv photometric o3 analyzer
Table of Contents

Advertisement

Servicing

Removing the Measurement Bench and Lowering the Partition Panel

Removing the
Measurement Bench
and Lowering the
Partition Panel
REMOVE SCREWS
Partition Panel
Loosen Captive Screw
Remove Screw
7-8
Model 49i Instruction Manual
The measurement bench can be removed and the partition panel can be
lowered to improve access to connectors and components. Refer to the
following steps when a procedure requires lowering the partition panel (see
Figure
7-3).
Remove Screw
Pass Through
Connectors
Figure 7-3. Removing the Measurement Bench and Lowering the Partition Panel
Equipment Required:
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF and unplug the power cord.
2. If the instrument is mounted in a rack, remove it from the rack.
Remove Screws
SLIDE BENCH OUT
LOWER PANEL
Remove Hidden Screw