Thermo Scientific 49i Instruction Manual page 279

Uv photometric o3 analyzer
Table of Contents

Advertisement

Thermo Fisher Scientific
set sample
set zero
set span
These commands set the current gas mode to sample, zero, or span. The
example that follows sets the instrument to span mode, that is, the
instrument is sampling span gas.
Send:
set span
Receive:
set span ok
gas unit
This command reports the current gas units (ppm or mg/m
reports that the gas unit is set to ppb.
Send:
gas unit
Receive:
gas unit ppb
set gas unit
3
unit = | ppm | mg/m
|
This command sets the gas units to ppm or mg/m
follows sets the gas units to mg/m
Send:
set gas unit mg/m3
Receive:
set gas unit mg/m3 ok
lamp
This reports the current photometer lamp setting. The example that follows
reports that the lamp setting is 72.9%
Send:
lamp setting
Receive:
lamp setting 72.9%
set lamp ddd.d
This command sets the photometer lamp setting to ddd.d, where ddd.d is a
floating-point number representing a percentage of the photometer lamp
setting. The example that follows sets the photometer lamp to 75.5%
Send:
set lamp 75.5
Receive:
set lamp 75.5 ok
lamp setting
This command reports the current photometer lamp setting. The example
that follows reports that the lamp setting is 72.9%.
Send:
lamp setting
Receive:
lamp setting 72.9%
C-Link Protocol Commands
Measurement Configuration
3
. The example that
3
.
Model 49i Instruction Manual
3
). The example
.
B-31