Servicing
Removing the Measurement Case Assembly and Lowering the Partition Panel
Removing the
Measurement Case
Assembly and Lowering
the Partition Panel
REMOVE SCREWS
Loosen
Partition Panel
Captive
Screw
Remove Screw
7-8
Model 48i Instruction Manual
The measurement case assembly can be removed and the partition panel
can be lowered to improve access to connectors and components. Refer to
the following steps when a procedure requires lowering the partition panel
(see
Figure
Pass Through
Connectors
Remove Screws
Figure 7–3. Removing the Measurement Case Assembly and Lowering the
Partition Panel
Equipment Required:
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Turn the instrument OFF and unplug the power cord.
2. If the instrument is mounted in a rack, remove it from the rack.
7–3).
Remove Screw
SLIDE CASE ASSEMBLY OUT
LOWER PANEL
Remove Hidden Screw