REMOVE SCREWS
Loosen
Captive
Screw
Thermo Fisher Scientific
Pass Through
Connectors
Partition Panel
Remove Screw
LOWER PANEL
Figure 7–3. Removing the Measurement Bench and Lowering the Partition
Panel
Equipment Required:
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF and unplug the power cord.
2. If the instrument is mounted in a rack, remove it from the rack.
3. Remove the cover.
4. Disconnect the plumbing connections at the rear of the measurement
bench.
Removing the Measurement Bench and Lowering the Partition Panel
Remove Screw
Remove Screws
SLIDE BENCH OUT
Remove Hidden Screw
▲
Model 43i Trace Level-Enhanced Instruction Manual
Servicing
7-7