Operation
Service Menu
RESTORE DEFAULT CAL A:
RANGE
Temperature Calibration
3-92
Model 49i Instruction Manual
•
In the Main Menu, choose Service > Select Flow Calibration A or B >
Set Defaults.
•
Press
•
Use
factory default values when pressed after
RESTORE
AVG
DIAGS ALARM
The Temperature calibration screen allows the user to view and set the
ambient temperature sensor calibration. The temperature calibration is
visible only when the instrument is in service mode. For more information
on the service mode, see "Service Mode" earlier in the chapter.
Note This adjustment should only be performed by an instrument service
technician.
•
In the Main Menu, choose Service > Temperature Calibration.
•
Use
•
Press
CALIBRATE AMBIENT TEMP:
CURRENTLY:
RANGE
to warn user and enable restore with
to overwrite pressure sensor calibration parameters with
RESTORE DEFAULT CAL A:
ARE YOU SURE YOU WANT
PRESS
,
,
and
to save set to value as actual value.
32.3
SET TO:
030.0
MOVE CURSOR
CHANGE VALUE
AVG
DIAGS ALARM
.
TO CONFIRM RESTORE
RANGE
AVG
DIAGS ALARM
to move and change the value.
o
C
o
C
SAVE
.
RESTORE
TO?x