Analog Current Output Range Values; Analog Voltage Output Range Values - Thermo Scientific 49i Instruction Manual

Uv photometric o3 analyzer
Table of Contents

Advertisement

C-Link Protocol Commands
I/O Configuration
B-40
Model 49i Instruction Manual
Send:
set analog iout range 4 1
Receive:
set analog iout range 4 1 ok
Table B-9. Analog Current Output Range Values
Range
1
2
0 [cannot be set to this, but may report]
analog vin channel
This command retrieves the analog voltage input channel data, both the
calculated value and the actual voltage. In the example that follows, the
"calculated" value of channel 1 is 75.325 degrees F, volts are 2.796. This
command responds with "feature not enabled" if the I/O expansion board is
not detected.
Send:
analog vin 1
Receive:
analog vin 1 75.325 2.796
analog vout range channel
This command reports the analog voltage output channel range, where
channel is between 1 and 6 inclusive, according to
Send:
analog vout range 2
Receive:
analog vout range 2 3
set analog vout range channel range
This command sets analog voltage output channel to the range, where
channel is between 1 and 6 inclusive, and range is set according to
Table
B-10. The example that follows sets channel 2 to the 0-10 V range.
Send:
set analog vout range 2 3
Receive:
set analog vout range 2 3 ok
Table B-10. Analog Voltage Output Range Values
Range
1
2
3
4
0 [cannot be set to this, but may report]
Output Range
0-20 mA
4-20 mA
Undefined
Table
B-10.
Output Range
0-1 V
0-100 mV
0-10 V
0-5 V
Undefined