Diagnostics Menu

DATE AND TIME:
19 MAR 2005
PRESS
RANGE
Diagnostics Menu

Program Version

3-64
Model 49i Instruction Manual
Use
and time.
Press
12:34:56
TO EDIT
AVG
DIAGS ALARM
The Diagnostics menu provides access to diagnostic information and
functions. This menu is useful when troubleshooting the instrument. The
analog input readings and analog input voltages are only displayed if the I/O
expansion board option is installed.
In the Main Menu, choose Diagnostics.
DIAGNOSTICS:
>PROGRAM VERSION
VOLTAGES
TEMPERATURES
PRESSURE
FLOWS
CELL A/B O3
INTENSITIES
RANGE
ANALOG INPUT READINGS
ANALOG INPUT VOLTAGES
DIGITAL INPUTS
RELAY STATES
TEST ANALOG OUTPUTS
INSTRUMENT CONFIGURATION
CONTACT INFORMATION
The Program Version screen (read only) shows the version number of the
program installed. Prior to contacting the factory with any questions
regarding the instrument, please note the program version number.
,
and
to move and change the value of the date
to edit and save the new date and time.
DATE AND TIME:
19 MAR 2005
SETTING: DAYS
RANGE
AVG
DIAGS ALARM
12:34:56
SET MONTHS
CHANGE VALUE
SAVE VALUE
AVG