C-Link Protocol Commands
Measurement Configuration
B-30
Model 49i Instruction Manual
custom range
range = | 1 | 2 | 3 |
This command reports the user-defined value of custom range 1, 2, or 3.
The example that follows reports that custom range 1 is defined to 55.0
ppm.
Send:
custom 1
Receive:
custom 1 5500E-2 ppm
set custom range value
set custom 1 value
set custom 2 value
set custom 3 value
set custom 1 range value
set custom 2 range value
set custom 3 range value
These commands are used to set the maximum concentration for any of the
three custom ranges 1, 2, or 3 to range value, where value is a floating-point
number representing concentration in ppm or mg/m
follows sets the custom 1 range to 55.5 ppm.
Send:
set custom 1 range 55.5
Receive:
set custom 1 range 55.5 ok
range mode
This command reports the current range mode.
Send:
range mode
Receive:
range mode single
set range mode mode
This command sets the current range mode to single, dual, or auto. The
example that follows sets the range mode to single.
Send:
set range mode single
Receive:
set range mode single ok
gas mode
This command reports the current mode of sample, zero, or span. The
example that follows reports that the gas mode is sample.
Send:
gas mode
Receive:
gas mode sample
3
. The example that
Thermo Fisher Scientific