RS-232 Connection

RS-485 Connection
Thermo Fisher Scientific
A null modem (crossed) cable is required when connecting the analyzer to
an IBM Compatible PC. However, a straight cable (one to one) may be
required when connecting the analyzer to other remote devices. As a general
rule, when the connector of the host remote device is female, a straight cable
is required and when the connector is male, a null modem cable is required.
Data Format:
1200, 2400, 4800, 9600, 19200, 38400, 57600, or 115200 BAUD
8 data bits
1 stop bit
no parity
All responses are terminated with a carriage return (hex 0D)
Refer to
Table 8-1
Table 8-1. RS-232 DB Connector Pin Configurations
DB9 Pin
Function
2
RX
3
TX
7
RTS
8
CTS
5
Ground
The instrument uses a four wire RS-485 configuration with automatic flow
control (SD). Refer to
Table 8-2. RS-485 DB Connector Pin Configuration
DB9 Pin
Function
2
+ receive
8
- receive
7
+ transmit
3
- transmit
5
ground
for the DB9 connector pin configuration.
Table 8-2
for the DB9 connector pin configuration.
System Description
I/O Components
Model 49i Instruction Manual
8-11