The partition panel of the measurement bench can be lowered to improve
access to connectors and components. Refer to the following steps when a
procedure requires lowering the partition panel (see
Pass Through
Connectors
Remove Screws
Figure 7–3. Removing the Measurement Bench and Lowering the Partition
Panel
Equipment Required:
Philips screwdriver
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF and unplug the power cord.
Removing the Measurement Bench and Lowering the Partition Panel
Remove Screw
Model 48i Trace Level-Enhanced Instruction Manual
Servicing
Figure
7–3).
SLIDE BENCH OUT
LOWER PANEL
Remove Hidden Screw