Thermo Scientific 49i Instruction Manual page 284

Uv photometric o3 analyzer
Table of Contents

Advertisement

C-Link Protocol Commands
Communications Configuration
B-36
Model 49i Instruction Manual
addr gw
This command reports the default TCP/IP gateway address.
Send:
addr gw
Receive:
addr gw 192.168.1.1
set addr gw address
This command sets the default gateway address, where address consists of
four numbers ranging from 0-255 inclusive, separated by ".".
Send:
set addr gw 192.168.1.1
Receive:
set addr gw 192.168.1.1 ok
addr ip
This command reports the IP address of the analyzer.
Send:
addr ip
Receive:
addr ip 192.168.1.15
set addr ip address
This command sets the analyzer's IP address, where address consists of four
numbers ranging from 0-255 inclusive, separated by ".".
Send:
set addr ip 192.168.1.15
Receive:
set addr ip 192.168.1.15 ok
addr nm
This command reports the TCP/IP netmask address.
Send:
addr nm
Receive:
addr nm 255.255.255.0
set addr nm address
This command sets the netmask address, where address consists of four
numbers ranging from 0-255 inclusive, separated by ".".
Send:
set addr nm 255.255.255.0
Receive:
set addr nm 255.255.255.0 ok
baud
This command reports the current baud rate for the serial port
(RS232/RS485). The example that follows reports that the current baud
rate is 9600.
Send:
baud
Receive:
baud 9600
Thermo Fisher Scientific