BENCH TEMPERATURE:
ACTUAL MIN:
SET MIN TO:
RANGE

Pressure

The Pressure screen displays the current pressure and sets the minimum and
maximum alarm limits. Acceptable alarm limits range from 200 to 1,000
mmHg. If the pressure reading goes beyond either the minimum or
maximum limit, an alarm is activated. The word "ALARM" appears in the
Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Pressure.
PRESSURE:
ACTUAL
>MIN
MAX
RANGE
The Minimum Pressure alarm limit screen is used to change the minimum
pressure alarm limit. The minimum and maximum pressure screens
function the same way.
In the Main Menu, choose Alarms > Pressure > Min or Max.
Use
Press
o
15.0
o
16.0
INC/DEC
SAVE VALUE
AVG
DIAGS ALARM
753.4 mmHg
200.0 mmHg
1000.0 mmHg
AVG
DIAGS ALARM
and
to increment or decrement the numeric value.
to save set to value as actual value.
C
C ?
Thermo Fisher Scientific