Servicing

Analog Output Calibration

Analog Output
Calibration
7-14
Model 49i Instruction Manual
Table 7-4. Analog Output Channels and Rear Panel Pin Connections
Voltage
Pin
Channel
1
14
2
33
3
15
4
34
5
17
6
36
Ground
16, 18, 19, 35, 37
Table 7-5. Analog Input Channels and Rear Panel Pin Connections
Input Channel
Pin
1
1
2
2
3
3
4
5
5
6
6
7
7
9
8
10
Ground
4, 8, 11, 14
Use the following procedure to calibrate the analog outputs if a meter
reading in the "Analog Output Testing" procedure differed by more than
one percent or after replacing the optional I/O expansion board.
Equipment Required:
Multimeter
1. Connect a meter to the channel to be adjusted and set to voltage or
current as appropriate.
Table 7-4
identifies the associated channels.
Current
Channel
1
2
3
4
5
6
Current Output Return
Figure 7-6
shows the analog output pins and
Pin
15
17
19
21
23
25
13, 16, 18, 20, 22, 24