Table B-8. Reply Termination Formats
Format
Reply Termination
00
<CR>
01
<NL> sum xxxx <CR>
where xxxx = 4 hexadecimal digits that represent the sum of all the
characters (bytes) in the message
host name
This command reports the host name string.
Send:
host name
Receive:
host name ISERIES
set host name string
This command sets the host name string, where string is 1-13 alphanumeric
characters.
Send:
set host name analyzer01
Receive:
set host name analyzer01 ok
instr name
This command reports the instrument name.
Send:
instr name
Receive:
instr name
o3 Analyzer
03 Analyzer
instrument id
This command reports the instrument id.
Send:
instrument id
Receive:
instrument id 49
set instrument id value
This command sets the instrument id to value, where value is a decimal
number between 0 and 127 inclusive.
Note Sending this command via RS-232 or RS-485 will require the host to
use the new id for subsequent commands.
Send:
set instrument id 50
Receive:
set instrument id 50 ok