Thermo Scientific 49i Instruction Manual page 102

Uv photometric o3 analyzer
Table of Contents

Advertisement

Operation
Diagnostics Menu
Interface Board Voltages
I/O Board Voltages
3-66
Model 49i Instruction Manual
MOTHERBOARD VOLTAGES:
3.3 SUPPLY
5.0 SUPPLY
15.0 SUPPLY
24.0 SUPPLY
-3.3 SUPPLY
RANGE
AVG
The Interface Board screen (read only) is used to display the current voltage
readings on the interface board.
In the Main Menu, choose Diagnostics > Voltages > Interface Board
Voltages.
INTERFACE BOARD VOLTAGES:
3.3 SUPPLY
5.0 SUPPLY
15.0 SUPPLY
-15.0 SUPPLY
24.0 SUPPLY
PHOTO LAMP
O3 LAMP
RANGE
AVG
The I/O Board screen (read only) is used to display the current voltage
readings on the I/O expansion board. This menu is only displayed if the I/O
expansion board option is installed.
In the Main Menu, choose Diagnostics > Voltages > I/O Board
Voltages.
I/O BOARD VOLTAGES:
3.3 SUPPLY
5.0 SUPPLY
24.0 SUPPLY
-3.3 SUPPLY
RANGE
AVG
3.3 V
5.0 V
15.0 V
24.0 V
-3.3 V
DIAGS ALARM
3.3 V
5.0 V
15.0 V
-15.0 V
24.0 V
9.6 V
17.2 V
DIAGS ALARM
3.3 V
5.0 V
24.0 V
-3.3 V
DIAGS ALARM
Thermo Fisher Scientific