Operation
Diagnostics Menu

Cell A/B O

Intensities

3-68
Model 49i Instruction Manual
FLOWS:
CELL A
CELL B
RANGE
AVG
The Cell A/B O
3
3
each cell, as well as the O
of the two cells).
In the Main Menu, choose Diagnostics > Cell A/B O
O3
PPB:
CELL A
CELL B
RANGE
AVG
The Intensities screen (read only) displays the current intensity in Cell A
and Cell B in Hertz. These intensities are read by detectors A and B,
respectively.
In the Main Menu, choose Diagnostics > Intensities.
INTENSITIES:
CELL A
CELL B
RANGE
AVG
0.608 LPM
0.612 LPM
DIAGS ALARM
screen (read only) displays the current O
reading displayed in the Run screen (the average
3
600
598
602
DIAGS ALARM
100425 Hz
100465 Hz
DIAGS ALARM
concentration in
3
.
3