I/O Configuration - Thermo Scientific 49i Instruction Manual

Uv photometric o3 analyzer
Table of Contents

Advertisement

I/O Configuration

Thermo Fisher Scientific
mode
This command reports what operating mode the instrument is in: local,
service, or remote. The example that follows shows that the instrument is in
the remote mode.
Send:
mode
Receive:
mode remote
set mode local
set mode remote
These commands set the instrument to local or remote mode. The example
that follows sets the instrument to the local mode.
Send:
set mode local
Receive:
set mode local ok
program no
This command reports the analyzer's model information and program
version number, which will be dependant on the current version.
Send:
program no
Receive:
program no iSeries 49i 01.00.01.074
set layout ack
This command disables the stale layout/layout change indicator ('*') that is
attached to each response if the layout has changed.
Send:
set layout ack
Receive:
set layout ack ok
analog iout range channel
This command reports the analog current output range setting for channels,
where channel must be between 1 and 6, inclusive. The example that follows
reports current output channel 4 to the 4-20 mA range, according to
Table
B-9. This command responds with "feature not enabled" if the I/O
expansion board is not detected.
Send:
analog iout range 4
Receive:
analog iout range 4 2
set analog iout range channel range
This command sets analog current output channel to the channel range
where channel is between 1 and 6 inclusive, and range is set according to
Table
B-9. The example that follows sets current output channel 4 to the
0-20 mA range. This command responds with "feature not enabled" if the
I/O expansion board is not detected.
C-Link Protocol Commands
I/O Configuration
Model 49i Instruction Manual
B-39