Optical Bench
Use the following to remove the optical bench assembly.
Assembly Removal
Equipment required:
Hex nut drive, 3/8 and 5/32 and hex drive, 5/32
1. Unplug all electrical as shown
(Figure
7–35).
Figure 7–35. Removing the Optical Bench Assembly
2. Using 3/8 hex nut drive, unfasten nut.
3. Using 5/32 hex drive, unfasten two #10-32 socket cap head screws.
4. Pull optical bench assembly upwards
(Figure
7–36).