Thermo Scientific 49i Instruction Manual page 92

Uv photometric o3 analyzer
Table of Contents

Advertisement

Operation
Instrument Controls Menu
Minimum and Maximum Value
3-56
Model 49i Instruction Manual
The MinimumValue screen is used to edit the zero (0) to full-scale (100)
value in percentages for the selected analog output channel. See Table 3-6
for a list of choices. The minimum and maximum output value screens
function the same way. The example below shows the set minimum value
screen.
In the Main Menu, choose Instrument Controls > IO Configuration >
Analog Output Config > Select Channel > Set Minimum or Maximum
Value.
Use
and
Press
to validate and save the new minimum value.
MINIMUM OUTPUT PERCENT:
SELECTED OUTPUT:
CURRENTLY:
SET TO:
CHANGE VALUE
RANGE
AVG
Table 3-6. Analog Output Zero to Full-Scale Table
Output
Zero % Value
O
Zero (0)
3
LO O
Zero (0)
3
HI O
Zero (0)
3
Range Status
Recommend not to change the setting for this output
Intensity A
User-set alarm min value
Intensity B
User-set alarm min value
Noise A
Zero (0)
Noise B
Zero (0)
Flow A
User-set alarm min value
Flow B
User-set alarm min value
Pressure
User-set alarm min value
Bench Temp
User-set alarm min value
Lamp Temp
User-set alarm min value
to increment or decrement the numeric value.
V ALL
N/A %
0000.5
% ?
SAVE
DIAGS ALARM
Full-Scale 100% Value
Range Setting
Range Setting
Range Setting
User-set alarm max value
User-set alarm max value
Range Setting
Range Setting
User-set alarm max value
User-set alarm max value
User-set alarm max value
User-set alarm max value
User-set alarm max value