Lamp Temperature

3
Thermo Fisher Scientific
ALARMS:
ALARMS DETECTED
>O3 LAMP TEMP
LAMP TEMP
BENCH TEMP
PRESSURE
FLOW A
FLOW B
RANGE
AVG
INTENSITY A
INTENSITY B
ZERO CHECK
SPAN CHECK
ZERO AUTOCAL
SPAN AUTOCAL
OZ LEVEL 1 CHECK
OZ LEVEL 2 CHECK
OZ LEVEL 3 CHECK
OZ LEVEL 4 CHECK
OZ LEVEL 5 CHECK
O3 CONCENTRATION
MOTHERBOARD STATUS
INTERFACE STATUS
I/O EXP STATUS
The O
Lamp Temperature screen (if the ozonator option is installed)
3
displays the current ozonator lamp temperature and sets the minimum and
maximum alarm limits. Acceptable alarm limits range from 65 to 75 °C. If
the O
lamp temperature reading goes beyond either the minimum or
3
maximum alarm limit, an alarm is activated. The word "ALARM" appears
in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > O
O3 LAMP TEMPERATURE:
ACTUAL
>MIN
MAX
RANGE
AVG
0
OK
OK
OK
OK
OK
OK
DIAGS ALARM
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
Lamp Temp.
3
o
68.8
C
o
60.0
C
o
80.0
C
DIAGS ALARM
Model 49i Instruction Manual
Operation
Alarms Menu
3-73