Thermo Scientific 49i Instruction Manual page 285

Uv photometric o3 analyzer
Table of Contents

Advertisement

Thermo Fisher Scientific
set baud rate
rate = | 1200 | 2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200 |
This command sets the instrument baud rate. The example that follows sets
the instrument's baud rate to 115200.
Note After the command is sent, the baud rate of the sending device must
be changed to agree with the instrument.
Send:
set baud 115200
Receive:
set baud 115200 ok
dhcp
This command reports the current state of use of DHCP on or off. DHCP
is used to assign an IP address to the analyzer automatically. The example
that follows shows that DHCP is on.
Send:
dhcp
Receive:
dhcp on
set dhcp onoff
These commands enables and disables the DHCP service by either on or off.
Changes to this parameter will only take effect when the analyzer is powered
up. The example that follows sets the DHCP service on.
Note When DHCP is set to on, the user-supplied addr gw, addr dns, addr
ip, and addr nm parameters are not used.
Send:
set dhcp on
Receive:
set dhcp on ok
format
This command reports the current reply termination format. The example
that follows shows that the reply format is 00, which means reply with no
checksum, according to
Send:
format
Receive:
format 00
set format format
This command sets the reply termination format, where format is set
according to
Table
B-8. The example that follows sets the reply termination
format to checksum.
Send:
set format 01
Receive:
set format 01 ok
Table
B-8.
C-Link Protocol Commands
Communications Configuration
Model 49i Instruction Manual
B-37

Advertisement

Table of Contents
loading

Table of Contents