Operation
Instrument Controls Menu
Temperature
Compensation

Pressure Compensation

3-30
Model 49i Instruction Manual
CYCLE TIME:
CURRENTLY:
SET TO:
RANGE
AVG
Temperature compensation provides compensation for any changes to the
instrument's output signal due to variations in sample gas temperature. The
Model 49i can be operated with or without temperature compensation.
When temperature compensation is on, the display shows the sample gas
temperature (measured by a thermistor on the Interface board). When
temperature compensation is off, the display shows the standard
temperature of 0.0 °C.
In the Main Menu, choose Instrument Controls > Temperature
Compensation.
Press
to toggle and set the temperature compensation on or off.

TEMPERATURE COMPENSATION:

COMP TEMP:
CURRENTLY:
SET TO:
RANGE
AVG
Pressure compensation provides compensation for any changes to the
instrument's output signal due to variations of sample pressure. The Model
49i can be operated with or without pressure compensation.
When pressure compensation is on, the first line of the display represents
the current sample pressure. When pressure compensation is off, the first
line of the display shows the standard pressure of 760 mmHg.
STANDARD
FAST ?
TOGGLE VALUE
DIAGS ALARM
o
0.0
C
OFF
ON ?
TOGGLE VALUE
DIAGS ALARM