Span Adjust; Additional Concentration Standards - Thermo Scientific 49i Instruction Manual

Uv photometric o3 analyzer
Table of Contents

Advertisement

Calibration
Calibration Procedure

Span Adjust

Additional Concentration
Standards
4-10
Model 49i Instruction Manual
If a strip chart recorder is used, it is recommended that it be adjusted to
obtain a record of zero drift and/or zero noise. This can be achieved by
using the zero offset capability of the recorder.
5. Record the stable zero air response as Z.
Use the following procedure to adjust span.
1. Generate an ozone concentration standard of approximately 80% of the
upper range limit (URL) of the ozone analyzer (such as, 0.4 or 0.8 ppm
for the 0.5 and 1.0 ppm ranges respectively).
2. Allow the instrument to sample this ozone concentration standard until
a stable response is obtained.
3. From the Main Menu choose Calibration > Calibrate Span. Use
and
to increment/decrement the known span gas concentration.
Press
to calibrate the instrument.
The recorder response will equal:
RecorderRe
sponse
(%Scale)
where:
URL = Upper range limit of the Model 49i, ppm
Z = Recorder response with zero air, % scale
[O
]
= Ozone concentration as determined by the calibration photometer, ppm
3
out
4. Record the ozone concentration as determined by the calibration
photometer and the corresponding analyzer response.
1. Generate several other ozone concentration standards (at least five others
are recommended) over the scale range of the instrument.
(O
)
3
out
100
Z
URL
(5)
Thermo Fisher Scientific