Thermo Scientific 48iQTL Instruction Manual
Thermo Scientific 48iQTL Instruction Manual

Thermo Scientific 48iQTL Instruction Manual

Trace level carbon monoxide analyzer

Advertisement

Quick Links

48iQTL
Instruction Manual
Trace Level Carbon Monoxide Analyzer
119014-00 • 27Sep2018

Advertisement

loading

Summary of Contents for Thermo Scientific 48iQTL

  • Page 1 48iQTL Instruction Manual Trace Level Carbon Monoxide Analyzer 119014-00 • 27Sep2018...
  • Page 3: Table Of Contents

    Contents Chapter 1 Introduction ......................1-1     iQ Series Instrument Platform............. 1-1   48iQTL Principle of Operation ............1-3   Specifications ..................1-5   Dimensions ..................1-6   Chapter 2 Installation and Setup ..................2-1     Unpacking and Inspection ..............2-1  ...
  • Page 4 Change Full Access Security Password ........3-107   USB Drive ................... 3-109   Firmware Update Via USB Drive ..........3-110   Download Data To USB Drive ..........3-111   Change USB Password .............. 3-114   User Contact Information ............3-116   48iQTL Instruction Manual Thermo Scientific...
  • Page 5 Cleaning the Optics ................5-6   IR Source Replacement ............... 5-8   Chapter 6   Troubleshooting ....................6-1   Safety Precautions ................6-1   Troubleshooting Guide ............... 6-1   Chapter 7 Servicing ......................7-1     Thermo Scientific 48iQTL Instruction Manual...
  • Page 6 Infrared Source ................. 8-3   CO Scrubber ..................8-3   Optical Bench DMC Board ............. 8-4   Instrument Temperature Control ............. 8-4   Fan (rear panel) ................8-4   Case Heater Assembly ..............8-4   48iQTL Instruction Manual Thermo Scientific...
  • Page 7 Safety and Equipment Damage Alerts ..........A-1   Warranty ..................... A-2   WEEE Compliance ................A-4   WEEE Symbol ................. A-4   Appendix B Quick Reference ....................B-1     List of Figures ..................B-1   List of Tables..................B-3 Thermo Scientific 48iQTL Instruction Manual...
  • Page 8 Contents Appendix C GNU Lesser General Public License ............C-1     GNU Lesser General Public License ........... C-1   viii 48iQTL Instruction Manual Thermo Scientific...
  • Page 9: Chapter 1 Introduction

    Dual range and auto range are standard features in the 48iQTL. Standard features also include an intelligent zero reference to optimize drift performance and dynamic filtering to maintain data stability and integrity in changing environmental conditions.
  • Page 10: Figure 1-1. 48Iqtl Front

    PC. The ePort control mirrors the same GUI look and feel as the instrument touchscreen providing a speedy and familiar operational experience. Figure 1–1. 48iQTL Front 48iQTL Instruction Manual Thermo Scientific...
  • Page 11: 48Iqtl Principle Of Operation

    1000 ppm. Gas sample is drawn into the 48iQTL through the rear panel of the analyzer and through the optical detection system using an embedded...
  • Page 12: Figure 1-2. 48Iqtl Flow Schematic

    Introduction 48iQTL Principle of Operation Figure 1–2. 48iQTL Flow Schematic Figure 1–3. 48iQTL Flow Schematic with Zero Span Manifold 48iQTL Instruction Manual Thermo Scientific...
  • Page 13: Specifications

    Introduction Specifications Specifications Table 1–1 lists the specifications for the 48iQTL. Table 1–1. 48iQTL Specifications Range 0-1000 (ppm or mg/m Zero Noise 0.01 ppm RMS Span Noise 0.5% of reading above 2.5 ppm Detection Limit 0.02 ppm Zero Drift <0.025 ppm (24 hour) Span Drift ±0.5% reading (24 hour)
  • Page 14: Dimensions

    Introduction Dimensions Dimensions Figure 1–4. Bench Mount Assembly (dimensions in inches [mm]) 48iQTL Instruction Manual Thermo Scientific...
  • Page 15: Figure 1-5. Rack Mount Assembly (Dimensions In Inches [Mm])

    Introduction Dimensions Figure 1–5. Rack Mount Assembly (dimensions in inches [mm]) Thermo Scientific 48iQTL Instruction Manual...
  • Page 16: Figure 1-6. Rack Mount Requirements

    Introduction Dimensions Figure 1–6. Rack Mount Requirements Figure 1–7. Rack Requirements Part 2 48iQTL Instruction Manual Thermo Scientific...
  • Page 17: Installation And Setup

    ▲ Unpacking and The 48iQTL is shipped complete in one container. If there is obvious damage to the shipping container when you receive the instrument, notify Inspection the carrier immediately and hold for inspection. The carrier is responsible for any damage incurred during shipment.
  • Page 18: Cover Removing And Replacing

    2. Press in both latches located on top cover and hold while pulling up to remove. Set upright. Figure 2–1. Removing the Cover 3. To replace, align cover and drop in. Latches will automatically snap in place. 48iQTL Instruction Manual Thermo Scientific...
  • Page 19: Mounting Options

    ● Bench Mount Positioned on bench, includes installing feet. See Figure 2–2. Equipment required: Slot drive, 5/16-inch 1. Fasten feet in position 1 or 2 to fit to the desired depth. Figure 2–2. Installing Feet Thermo Scientific 48iQTL Instruction Manual...
  • Page 20: Rack Mount

    4. Use the same four 8-32 pan head screws to secure it. 5. Install the handles with the four 8-32 flat head screws that came with the handle kit on the backside. See Figure 2–4. 48iQTL Instruction Manual Thermo Scientific...
  • Page 21: Figure 2-4. Installing Ears And Handles

    Installation and Setup Mounting Options Figure 2–4. Installing Ears and Handles Thermo Scientific 48iQTL Instruction Manual...
  • Page 22: Setup Procedure

    “Connecting External Devices” on page Communications > “Analog I/O” on page 3-50, and “Digital I/O” on page 3-52. 6. Plug the instrument into an outlet of the appropriate voltage and frequency. 48iQTL Instruction Manual Thermo Scientific...
  • Page 23: Figure 2-5. 48Iqtl Rear Panel

    Installation and Setup Setup Procedure Icon Here The 48iQTL is supplied with a three-wire grounding cord. Under no circumstances should this grounding system be defeated. ▲ Figure 2–5. 48iQTL Rear Panel Figure 2–6. Atmospheric Dump Bypass Plumbing Thermo Scientific 48iQTL Instruction Manual...
  • Page 24: Startup

    For more information about instrument parameters, see the “Operation” chapter. 4. Before beginning the actual monitoring, perform a multipoint calibration as described in the “Calibration” chapter. Figure 2–7. Front Panel and Touchscreen Display 48iQTL Instruction Manual Thermo Scientific...
  • Page 25: Chapter 3 Operation

    The Status Bar, located at the bottom, includes the Back button, Access Levels, Health Check, Favorites, Date and Time, and Contact Information. Home Screen (single range mode) Title Bar User Interface Status Bar Thermo Scientific 48iQTL Instruction Manual 3-1...
  • Page 26 Operation Instrument Display Home Screen (dual or auto range mode) Title Bar User Interface Status Bar 48iQTL Instruction Manual Thermo Scientific...
  • Page 27 The user will be directed to the favorites screen where the user chooses the button position. To remove a favorite button from the favorites screen, press and hold button for 2 seconds. Clock: Displays current date and time. ● Thermo Scientific 48iQTL Instruction Manual...
  • Page 28 Operation Instrument Display Thermo Scientific Information button: Shows contact information. ● 48iQTL Instruction Manual Thermo Scientific...
  • Page 29: Main Menus And Keypads

     Compensation  Pressure Calibration  Scrubber Efficiency  Intelligent Zero  Reference Communications Instrument Settings  Display Setup  Alarm Setpoints  Language  Clock  Pump Power  Reboot Instrument Configuration Security Access Levels USB Drive  Firmware Update Via USB  Drive  Download Data to USB Drive  Change USB Password User Contact Information Update Bootloader Thermo Scientific 48iQTL Instruction Manual...
  • Page 30: Numeric Keypad

    Enter a new value using the keypad, and then select the Enter button to set the new value or press the Cancel button to exit the keypad screen and return to the previous screen without saving the value. 48iQTL Instruction Manual Thermo Scientific...
  • Page 31: Alphanumeric Keypad

    Cancel button to exit the keypad screen and return to the previous screen without saving the value. The alphanumeric keypad is only available when the user needs to enter alphabet characters. Thermo Scientific 48iQTL Instruction Manual...
  • Page 32 Operation Instrument Display 48iQTL Instruction Manual Thermo Scientific...
  • Page 33: Calibration

    Home Screen>Calibration (dual or auto range mode) The Calibration menu contains the following information: Calibrate Background: Sets the CO reading to zero. ● Calibrate Span Coefficient: Sets the span coefficient when in single ● range mode. Thermo Scientific 48iQTL Instruction Manual...
  • Page 34 (internal zero) or an external source of zero air (through the zero port, external zero), when in “zero” gas mode. Advanced Calibration: Calibrates the instrument using a manual ● zero/span calibration, a multipoint calibration and provides calibration history. 3-10 48iQTL Instruction Manual Thermo Scientific...
  • Page 35: Calibrate Background

    Calculated Background: Read only. Displays what the current user-set ● background will become when the calibrate button is pressed. Calibrate: When pressed, updates the background, making the ● concentration go to zero. Thermo Scientific 48iQTL Instruction Manual 3-11...
  • Page 36: Calibrate Span Coefficient

    Before making an adjustment, be sure the analyzer samples span gas for at least 6–7 minutes. Home Screen>Calibration>Calibrate Span Coefficient (single range mode) Home Screen>Calibration>Calibrate High Range Span Coefficient (dual or auto range mode) 3-12 48iQTL Instruction Manual Thermo Scientific...
  • Page 37 Range Span Concentration” value is entered, the new calculated low range span coefficient is displayed when in dual or auto range mode. Calibrate: When pressed, updates the coefficient and the concentration ● should match the span concentration. Thermo Scientific 48iQTL Instruction Manual 3-13...
  • Page 38: Zero/Span Schedule

    Zero Duration: Sets how long zero air is sampled by the instrument. ● Span Duration: Sets how long span gas is sampled by the instrument. ● Purge Duration: Sets how long the purge period will be at the end of ● the schedule. 3-14 48iQTL Instruction Manual Thermo Scientific...
  • Page 39 1:1. (This means that each time the schedule is run, both the zero duration and span duration occurs.) The zero/span ratio is allowable between 1 to 99. If 99 is chosen, the schedule should only perform the Span on the 99th iteration. Thermo Scientific 48iQTL Instruction Manual 3-15...
  • Page 40: Advanced Calibration

    Manual Calibration: The user manually adjusts the background or span ● coefficient. Multipoint Calibration: Up to three gas concentrations (cal-points) may ● be calibrated. Three cal-points will give the most accurate readings over the entire range. 3-16 48iQTL Instruction Manual Thermo Scientific...
  • Page 41 Low Range Multipoint Calibration: Up to three gas concentrations (cal- ● points) for low range may be calibrated. Three cal-points will give the most accurate readings over the entire range. Calibration History: Lists all calibrations performed and calibration ● checks. Thermo Scientific 48iQTL Instruction Manual 3-17...
  • Page 42: Manual Calibration

    The Manual Calibration menu contains the following information: Adjust Background: Allows the user to manually adjust the zero ● background. Adjust Span Coefficient: Allows the user to manually adjust the span ● coefficient when in single range mode. 3-18 48iQTL Instruction Manual Thermo Scientific...
  • Page 43 Adjust Low Range Span Coefficient: Allows the user to manually adjust ● the low range span coefficient when in dual or auto range mode. Reset Background to 0.000 and Span Coefficient to 1.000: Resets all ● backgrounds and coefficients. Thermo Scientific 48iQTL Instruction Manual 3-19...
  • Page 44 Adjust Background: User manually adjusts zero background. ● Adjusted Concentration: Read only. Shows adjusted concentration based ● on adjusted zero background. Calibrate: Calibrates zero background by saving the newly adjusted zero ● background value. 3-20 48iQTL Instruction Manual Thermo Scientific...
  • Page 45 Adjust Span Coefficient: User manually adjusts span coefficient when in ● single range mode. Adjust High Range Span Coefficient: User manually adjusts high range ● span coefficient when in dual or auto range mode. Thermo Scientific 48iQTL Instruction Manual 3-21...
  • Page 46 Adjusted Low Range Concentration: Read only. Shows adjusted low ● range concentration based on adjusted low range span coefficient when in dual or auto range mode. Calibrate: Calibrates span coefficient by saving the newly adjusted span ● coefficient. 3-22 48iQTL Instruction Manual Thermo Scientific...
  • Page 47: Multipoint Calibration

    Point 3: Third cal point used in the calibrated multipoint polynomial ● curve. First Coefficient: Read only. Value of the first coefficient in the ● polynomial curve. Second Coefficient: Read only. Value of the second coefficient in the ● polynomial curve. Thermo Scientific 48iQTL Instruction Manual 3-23...
  • Page 48 Operation Calibration Third Coefficient: Read only. Value of the third coefficient in the ● polynomial curve. Reset Points to Default Values: Resets multipoint coefficients to default ● values. 3-24 48iQTL Instruction Manual Thermo Scientific...
  • Page 49 Point 1 Coefficient: Read only. After selecting the value on the Edit ● Point 1 Span Concentration button and after pressing the Calibrate button, the Point 1 Coefficient value is updated. Calibrate: Calibrates point 1 coefficient. ● Thermo Scientific 48iQTL Instruction Manual 3-25...
  • Page 50: Calibration History

    Event: Lists the type of calibration event. ● Result: Concentration result. ● Target: Concentration setpoint value. ● Units: Displays units for each item. ● Average Time: Averaging time used during the calibration or calibration ● check. 3-26 48iQTL Instruction Manual Thermo Scientific...
  • Page 51: Data

    View Data Log (User Defined Time): User selects the start and end time ● for viewing the data. Table shows most recent data on top. Advanced Data Setup: Allows the user to set up the parameters of how ● the data is stored and streamed. Thermo Scientific 48iQTL Instruction Manual 3-27...
  • Page 52: View Data Log (Last Hour)

    ▲ The View Data Log (Last Hour) screen contains the following options: Graph: Displays data graph for the column selected. The graph time ● axis is defined by the data set in the table. 3-28 48iQTL Instruction Manual Thermo Scientific...
  • Page 53: View Data Log (Last 24 Hours)

    The View Data Log (Last 24 Hours) screen contains the following options: Graph: Displays data graph for the column selected. The graph time ● axis is defined by the data set in the table. Thermo Scientific 48iQTL Instruction Manual 3-29...
  • Page 54: View Data Log (User Defined Time)

    Minutes: Sets minutes of data logging start time. ● Save Data Logging Start Time: Pressing this button saves the start time ● and follows directly to the end time selection for the data logging screen. 3-30 48iQTL Instruction Manual Thermo Scientific...
  • Page 55 Time button saves the end time and follows directly to the data logging table. Note End time should not be greater than 1 year from start time . ▲ Note The datalogging table is limited to 10,000 points. ▲ Thermo Scientific 48iQTL Instruction Manual 3-31...
  • Page 56: Advanced Data Setup

    The Advanced Data Setup screen contains the following information: Data Logging Setup: User selects the parameters for collecting logged ● data. Streaming Data Setup: User selects the parameters for streaming data to ● a computer in real time. 3-32 48iQTL Instruction Manual Thermo Scientific...
  • Page 57: Data Logging Setup

    When set to minimum or maximum, the minimum or maximum value during the period will be recorded. Erase Data Log Records: Allows the user to erase all values in the data log ● and updates the data logging table. Thermo Scientific 48iQTL Instruction Manual 3-33...
  • Page 58: Select Data Logging Variables

    Use the  and  buttons to scroll through the variables. Select the variables to log by pressing the corresponding cells. Next, press the Commit Changes button to save selections. Yellow buttons indicate that the variable is selected. Home Screen>Data>Advanced>Data Logging Setup>Select Data Logging Variables 3-34 48iQTL Instruction Manual Thermo Scientific...
  • Page 59: Streaming Data Setup

    Show Labels: Toggles on/off. When on, shows variable labels to the left ● of the variable values. Show Timestamp: Toggles on/off. When on, shows timestamp at the ● beginning of each row of data. Thermo Scientific 48iQTL Instruction Manual 3-35...
  • Page 60: Select Streaming Variables

    Use the  and  buttons to scroll through the variables. Select the variables to log by pressing the corresponding cells. Next, press the Commit Changes button to save selections. Yellow buttons indicate that the variable is selected. Home Screen>Data>Advanced>Streaming Data Setup>Select Streaming Variables 3-36 48iQTL Instruction Manual Thermo Scientific...
  • Page 61: Settings

    USB Drive: User can update instrument firmware, download data, and ● change USB password. User Contact Information: User sets up their contact information. ● Update Bootloader: Used to update bootloader when an update to the ● bootloader is available. Thermo Scientific 48iQTL Instruction Manual 3-37...
  • Page 62: Health Check

    The iQ360 feature is a paid subscription enabling or disabling ● the instrument to send automated emails to technical support when an alarm or alert is triggered. Firmware Version: Shows the instrument’s firmware version. ● 3-38 48iQTL Instruction Manual Thermo Scientific...
  • Page 63: Status And Alarms

    Valve and Pump Resets: User can reset valve and pump power. ● Analog Input/Output: Displays analog input/output alarms and faults. ● Digital Input/Output: Displays digital input/output alarms and faults. ● Serial Numbers: Displays all the serial numbers for the instrument. ● Thermo Scientific 48iQTL Instruction Manual 3-39...
  • Page 64: Concentration

    High alarm shows user defined limit for acceptable background check offset. Span Check Offset: Displays concentration based on the last ● attempted span calibration. High alarm shows user defined limit for 3-40 48iQTL Instruction Manual Thermo Scientific...
  • Page 65 High alarm shows the user defined limit for acceptable intelligent zero reference offset. Note If both the low alarm and high alarms are set to zero, then no alarm will show. ▲ Thermo Scientific 48iQTL Instruction Manual 3-41...
  • Page 66: Ndir Bench

    Sample Intensity: Displays the current sample intensity reading. ● Reference Intensity: Displays the current reference intensity reading. ● Bench Pressure: Displays the current bench pressure reading. ● Flow: Displays the current sample flow reading. ● 3-42 48iQTL Instruction Manual Thermo Scientific...
  • Page 67 Operation Settings Instrument Temperature: Displays the current instrument ● temperature reading. In the 48iQTL, the instrument temperature is ° controlled at 43 Bench Temperature: Displays the current bench temperature ● reading. User can adjust low and high alarm limits. Bench Temperature too Low: Displays OK/Fail for bench ●...
  • Page 68 24V Diagnostic: Displays current voltage readings. Alarm limits ● are not changeable. -15V Diagnostic: Displays current voltage readings. Alarm limits ● are not changeable. Bias Voltage: Displays current voltage readings. Alarm limits are ● not changeable. 3-44 48iQTL Instruction Manual Thermo Scientific...
  • Page 69: Flow And Pressure

    Pump Pressure: Displays the current pump pressure reading. ● Instrument Temperature: Displays the current instrument ● temperature reading. User can adjust low and high alarm limits. . ° In the 48iQTL, the instrument temperature is controlled at 43 Thermo Scientific 48iQTL Instruction Manual 3-45...
  • Page 70 3.3V Diagnostic: Displays current voltage readings. Alarm limits ● are not changeable. 5.0V Diagnostic: Displays current voltage readings. Alarm limits ● are not changeable. 24V Diagnostic: Displays current voltage readings. Alarm limits ● are not changeable. 3-46 48iQTL Instruction Manual Thermo Scientific...
  • Page 71: Peripherals Support

    Zero Valve: Displays whether or not the zero valve is activated. ● Span Valve: Displays whether or not the span valve is activated. ● Pump: Displays whether or not the pump is activated. ● Thermo Scientific 48iQTL Instruction Manual 3-47...
  • Page 72 24V Step Board 1: Displays OK/Fail for step board 1. ● 5.0V Step Board 2: Displays OK/Fail for step board 2 (if ● installed). 24V Step Board 2: Displays OK/Fail for step board 2 (if ● installed). 3-48 48iQTL Instruction Manual Thermo Scientific...
  • Page 73: Valve And Pump Resets

    The Valve and Pump Resets screen contains the following information: Sample Valve Reset: Resets sample valve. ● Zero Valve Reset: Resets zero valve. ● Span Valve Reset: Resets span valve. ● Pump Reset: Resets pump. ● Sample/Zero Valve Reset: Resets sample/zero valve. ● Thermo Scientific 48iQTL Instruction Manual 3-49...
  • Page 74 Current Output Channel 1–6: Displays real-time current output for ● each channel. Chip Temperatures: Displays OK/Fail for chip temperatures. ● Chip 1–3 Communication: Displays OK/Fail for each chip ● communication. Test Mode: Displays test mode on or off. ● 3-50 48iQTL Instruction Manual Thermo Scientific...
  • Page 75 5.0V Ref Diagnostic: Displays current voltage readings. Alarm ● limits are not changeable. 15V Diagnostic: Displays current voltage readings. Alarm limits ● are not changeable. -15V Diagnostic: Displays current voltage readings. Alarm limits ● are not changeable. Thermo Scientific 48iQTL Instruction Manual 3-51...
  • Page 76 External Alarm 1–3: Displays OK/Fail for external alarms. ● Relay Test Mode: Displays relay test mode on or off. ● Solenoid Test Mode: Displays solenoid test mode on or off. ● Board Communication: Displays OK/Fail for communication status. ● 3-52 48iQTL Instruction Manual Thermo Scientific...
  • Page 77 3.3V Diagnostic: Displays current voltage readings. Alarm limits ● are not changeable. 5.0V Diagnostic: Displays current voltage readings. Alarm limits ● are not changeable. 24V Diagnostic: Displays current voltage readings. Alarm limits ● are not changeable. Thermo Scientific 48iQTL Instruction Manual 3-53...
  • Page 78: Serial Numbers

    NDIR Bench: NDIR bench serial number. ● Flow and Pressure: Flow and pressure serial number. ● Peripherals Support: Peripherals support serial number. ● Analog I/O: Analog I/O serial number. ● Digital I/O: Digital I/O serial number. ● 3-54 48iQTL Instruction Manual Thermo Scientific...
  • Page 79: Predictive Diagnostics

    The Predictive Diagnostics screen contains the following information: Filter Wheel ● Sample Pump ● Capillary ● Flow Path ● IR Source ● Sample Valve ● Zero Valve ● Span Valve ● Sample/Zero Valve ● Thermo Scientific 48iQTL Instruction Manual 3-55...
  • Page 80: Maintenance

    Preventive Maintenance: Shows suggested service interval and time left ● for component replacement. Change Part: User logs component fix. ● Maintenance History: Shows the log of all recorded component fixes. ● Advanced Maintenance: Resets all preventive maintenance items. ● 3-56 48iQTL Instruction Manual Thermo Scientific...
  • Page 81: Preventive Maintenance

    ● Reset button and the “Months Left” value resets to the “Service Interval in Months” value. Alert: Allows the user to opt out of receiving preventive ● maintenance alerts. Displays Enabled/Disabled for each component. Thermo Scientific 48iQTL Instruction Manual 3-57...
  • Page 82 Charcoal: Service interval for charcoal. ● Dri-Rite: Service interval for dri-rite. ● Scrubber: Service interval for scrubber. ● Scrubber Efficiency Check: Service interval for scrubber efficiency ● check. Permeation Dryer: Service interval for permeation dryer. ● 3-58 48iQTL Instruction Manual Thermo Scientific...
  • Page 83: Change Part

    Fix: User chooses from new, rebuilt, cleaned, and unknown. ● Comment: User can write a brief comment, which will be saved to the ● preventive maintenance history table. Commit: User commits and saves the selected part fix. ● Thermo Scientific 48iQTL Instruction Manual 3-59...
  • Page 84: Maintenance History

    Fix: The type of maintenance. ● Date: Shows date/time when service was logged. ● Service Months: Amount of time in months since last service. ● Comments: Shows comments entered from time of change. ● 3-60 48iQTL Instruction Manual Thermo Scientific...
  • Page 85: File Sharing And Support

    Email Health Check Report to Personal Account: Sends the health report ● file to a personal account via email. Request a Field Service Visit: Sends a field service visit request to ● technical support. Thermo Scientific 48iQTL Instruction Manual 3-61...
  • Page 86: Iq360

    Status and Renewal Request: Allows the user to access the iQ360 menu ● once the subscription has been enabled. Cancel and Return to the Home Screen: Exits screen and returns to the ● Home Screen. 3-62 48iQTL Instruction Manual Thermo Scientific...
  • Page 87 Enter Subscription Access Code: Allows the user to enter the subscription ● passcode to enable an iQ360 subscription. Exit and Return to the Home Screen: Exits screen and returns to the ● Home Screen. Thermo Scientific 48iQTL Instruction Manual 3-63...
  • Page 88 Subscription Expiration Date: Informs the user when the ● subscription will expire. iQ360 Test Connection: Allows the user to verify the network ● connection is working. iQ360 Subscription Renewal: Allows the user to request a subscription ● renewal via email. 3-64 48iQTL Instruction Manual Thermo Scientific...
  • Page 89: Measurement Settings

    Range Mode Selection: User can choose between the various range ● modes: single, dual, or auto. For more information, see “Range Mode Selection” on page 3-69 Range Setting: Sets the concentration range for the analog outputs when ● in single range mode. Thermo Scientific 48iQTL Instruction Manual 3-65...
  • Page 90 Gas Units: Defines the units in which CO concentration reading is ● expressed. Dilution Ratio: Serves as a multiplier when dilution gas is utilized. ● Advanced Measurement Settings: Advanced settings affecting CO ● readings. 3-66 48iQTL Instruction Manual Thermo Scientific...
  • Page 91: Averaging Time

    Settings Averaging Time The Averaging Time screen allows the user to choose dynamic filtering or a manually selected (static) averaging time. The 48iQTL, by default, uses dynamic filtering to smooth output data. Dynamic Filtering allows for data smoothing without compromising response time.
  • Page 92 High Range Averaging Time: Sets high averaging time when in dual or ● auto range mode and when dynamic filtering is disabled. Low Range Averaging Time: Sets low averaging time when in dual or ● auto range mode and when dynamic filtering is disabled. 3-68 48iQTL Instruction Manual Thermo Scientific...
  • Page 93: Range Mode Selection

    100 ppm. Sample concentrations below 50 ppm are output based on low range selection and sample concentrations above 50 ppm are output based on high range selection. When the low range Thermo Scientific 48iQTL Instruction Manual 3-69...
  • Page 94 ▲   3-70 48iQTL Instruction Manual Thermo Scientific...
  • Page 95: Range Setting

    High Range Setting or Low Range Setting, brings up a numeric keypad whereby the user can select a range. Settable ranges according to unit selection include: 501000000 ppb 11000 ppm 5e-60.1 % μg/m 501000000 μg/m mg/m 11000 mg/m 0.0011 g/m Thermo Scientific 48iQTL Instruction Manual 3-71...
  • Page 96: Gas Mode

    Zero: Used when calibrating the background of the instrument. When ● pressed, sets the instrument to zero mode. Span: Used when calibrating the span coefficient. When pressed, sets ● the instrument to span mode. 3-72 48iQTL Instruction Manual Thermo Scientific...
  • Page 97: Gas Units

    The Gas Units screen contains the following information: ppb: parts per billion. ● ppm: parts per million. ● %: percent. ● μg/m3: micrograms per meter cubed. ● mg/m3: milligrams per meter cubed. ● g/m3: grams per meter cubed. ● Thermo Scientific 48iQTL Instruction Manual 3-73...
  • Page 98: Advanced Measurement Settings

    Scrubber Efficiency: Allows the user to conduct the scrubber efficiency ● test. Intelligent Zero Reference: Allows the user to enable/disable the ● intelligent zero reference (i0 reference) feature. This feature is enabled, by default. 3-74 48iQTL Instruction Manual Thermo Scientific...
  • Page 99 Calibration The Detector Gain Calibration screen contains the following information: Across: ● Description: Defines the actions the user can do. ● Detector Gain: User can manually set or automatically calibrate the ● detector gain. Thermo Scientific 48iQTL Instruction Manual 3-75...
  • Page 100 Down: ● Manual Entry: Shows current value. If the initial S/R cell is pressed, ● user can manually set the initial S/R value. The instrument should be sampling zero gas when changing the S/R. 3-76 48iQTL Instruction Manual Thermo Scientific...
  • Page 101 User can Stop calibration by pressing the Stop button. End Cal: When pressed, the auto calibration is interrupted and the ● value does not change. Default Values: When pressed, the default S/R value is saved. ● Thermo Scientific 48iQTL Instruction Manual 3-77...
  • Page 102 Standard Temperature for Gas Units: Allows the user to enter a value ● between 0 and 30 °C to correct the g/m , mg/m , and g/m concentration calculations. 3-78 48iQTL Instruction Manual Thermo Scientific...
  • Page 103 Under the column labeled reading, the user enters the pressure in mmHg. Under the column labeled Calibration, the user presses the Start button to calibrate the low point of the sensor. Reset all values: Resets values to default. ● Thermo Scientific 48iQTL Instruction Manual 3-79...
  • Page 104 Settings>Scrubber Efficiency>Continue to Scrubber Efficiency Test The Scrubber Efficiency screen contains the following information: Across: ● Description: Lists items in table. ● Scrubber Efficiency: User can see concentrations/calculated scrubber ● efficiency and initiate/stop the automatic scrubber efficiency test. 3-80 48iQTL Instruction Manual Thermo Scientific...
  • Page 105 The user can stop the test pressing the Stop button. End Cal: When Stop is pressed, the scrubber efficiency test is ● interrupted. Last Test Completed: Displays the date of the last successfully ● completed scrubber efficiency test. Thermo Scientific 48iQTL Instruction Manual 3-81...
  • Page 106 Settings Intelligent Zero The Intelligent Zero Reference (i0 Reference) feature is enabled, by default, on the 48iQTL, and the button can be toggled to enable/disable the Reference feature. In essence, the i0 Reference feature is implemented to allow for the...
  • Page 107: Communications

    ID number if two or more of instruments of the same model are connected to one computer. Valid instrument ID numbers are from 0 to 127. The 48iQTL has a default instrument ID of 1. Bayern Hessen Settings: User selects the registers (measured values) that ●...
  • Page 108 Operation Settings Note Refer to the “Bayern Hessen Protocol” section in the Communications Manual for a detailed description. ▲ 3-84 48iQTL Instruction Manual Thermo Scientific...
  • Page 109: Wired Tcp/Dhcp

    The Wired TCP/DHCP screen contains the following information: DHCP: Toggles DHCP enabled/disabled. ● Dynamic IP Address: Dynamic IP address of the instrument. ● Dynamic Netmask: Dynamic Netmask of instrument. ● Dynamic Gateway: Dynamic Gateway of instrument. ● Thermo Scientific 48iQTL Instruction Manual 3-85...
  • Page 110 ● DHCP is disabled. DNS Server Address: DNS IP address of instrument. This is settable ● when DHCP is disabled. Wired MAC Address: Instrument MAC address. ● Host Name: Host name of instrument. ● 3-86 48iQTL Instruction Manual Thermo Scientific...
  • Page 111: Serial Rs-232/485

    Bits: User selectable between 7 and 8. ● Parity: User selectable between None, Even, and Odd. ● Stop Bits: User selectable between 1 and 2. ● RS 232/485: User selectable between RS-232 and RS-485. ● Thermo Scientific 48iQTL Instruction Manual 3-87...
  • Page 112: Analog I/O

    Analog Out Calibration: Allows the user to calibrate the analog out ● voltage and current to account for varying loads attached to the terminals. 3-88 48iQTL Instruction Manual Thermo Scientific...
  • Page 113: Digital I/O

    Digital Out (Relays): Allows the user to view relay outputs. ● Digital Out (Solenoids): Allows the user to view solenoid outputs. ● Advanced Digital I/O: Allows user to test the digital out relays and ● solenoids. Thermo Scientific 48iQTL Instruction Manual 3-89...
  • Page 114: Email Server (Smtp)

    From Email Address: The email address that goes in the From field in ● emails. SMTP Server Port: Server port of user’s email server. ● Email Password: Password for SMTP server. ● Email UserName: User name that is authorized to send email through ● SMTP server. 3-90 48iQTL Instruction Manual Thermo Scientific...
  • Page 115: Bayern Hessen Settings

    Home Screen>Settings>Communications>Bayern Hessen Settings The Bayern Hessen Settings screen contains the following information: Register 1–8: Sets the measured value logged in the respective register ● location, in response to a valid query (DA) command. Thermo Scientific 48iQTL Instruction Manual 3-91...
  • Page 116 Use the  and  buttons to scroll through the registers. Select the variable to log by pressing the corresponding cell. The list is based on the instrument’s configuration. Home Screen>Settings>Communications>Bayern Hessen Settings>Register 1–8 3-92 48iQTL Instruction Manual Thermo Scientific...
  • Page 117: Instrument Settings

    Alarm Setpoints: View and set all available alarm minimum and ● maximum values. Language: Read only. ● Clock: Sets date and time. ● Pump Power: Manually enables/disables the pump. ● Reboot Instrument: Reboots the instrument. ● Thermo Scientific 48iQTL Instruction Manual 3-93...
  • Page 118: Display Setup

    Power Save: Minutes before screen times out. Toggles enable/disable. ● Power Save Setting: Option whereby the user can display a black screen ● after a set amount of inactivity. Brightness: Sets the brightness of the display. ● 3-94 48iQTL Instruction Manual Thermo Scientific...
  • Page 119: Alarm Setpoints

    ● Down: ● CO: CO concentration alarm. ● Bkg Check Offset: User can set the maximum allowable background ● reading offset for calibration and calibration checks. This is set with the high alarm only. Thermo Scientific 48iQTL Instruction Manual 3-95...
  • Page 120 Span Check Offset: User can set the maximum allowable span ● reading offset for calibration and calibration checks. This is set with the high alarm only. Instrument Temperature: Instrument temperature alarm. ● Bench Pressure: Bench pressure alarm. ● Flow: Flow pressure alarm. ● 3-96 48iQTL Instruction Manual Thermo Scientific...
  • Page 121: Clock

    Month ● ● Year ● Hours ● Minutes ● Seconds ● Date / Time Parameters: User chooses time zone, timer server, and date ● format. Commit: When pressed, date and time are saved. ● Thermo Scientific 48iQTL Instruction Manual 3-97...
  • Page 122: Date / Time Parameters

    The Date / Time Parameters screen contains the following information: Time Zone: User selects time zone from table. ● Time Server Enabled/Disabled: User can enabled/disable the time server ● to get periodic clock updates. Date Format: User selects date format. ● 3-98 48iQTL Instruction Manual Thermo Scientific...
  • Page 123: Time Zone

    ● Mountain Time Zone(UTC-7) ● Mountain Daylight Savings(UTC-6) ● Central Time Zone(UTC-6) ● Central Daylight Savings((UTC-5) ● Eastern Time Zone(UTC-5) ● Eastern Daylight Savings(UTC-4) ● Atlantic Time Zone(UTC-4) ● Mid-Atlantic(UTC-3) ● South Georgia(UTC-2) ● Thermo Scientific 48iQTL Instruction Manual 3-99...
  • Page 124 ● Gulf Standard Time(UTC+4) ● Yekaterinburg Time(UTC+5 ● Omsk Time(UTC+6) ● Indochina Time(UTC+7) ● ASEAN Common Time(UTC+8) ● Japan Standard Time(UTC+9) ● Chamorro Time Zone(UTC+10) ● Sredmnekolymsk Time(UTC+11) ● New Zealand Standard Time(UTC+12) ● 3-100 48iQTL Instruction Manual Thermo Scientific...
  • Page 125: Time Server

    Time Server: Enables/Disables periodic clock updates from an NTP ● (Network Time Protocol) source. Set Time Server: User can choose specific time server. ● Set Default: When pressed, default time server will be used. ● Thermo Scientific 48iQTL Instruction Manual 3-101...
  • Page 126: Date Format

    Use the  and  buttons to select. Home Screen>Settings>Instrument Settings>Clock>Date / Time Parameters>Date Format The Date Format screen contains the following information: U.S. Format mm/dd/yyyy ● European Format dd/mm/yyyy ● ISO 8601 yyyy-mm-dd ● 3-102 48iQTL Instruction Manual Thermo Scientific...
  • Page 127: Configuration

    Zero/Span Assembly: Enables zero/span option. ● Predictive Diagnostics: Enables predictive diagnostics option. ● Communications Board: Enables RS-232 or RS-485 communication ● board option. Analog I/O: Enables analog I/O option. ● Digital I/O: Enables digital I/O option. ● Thermo Scientific 48iQTL Instruction Manual 3-103...
  • Page 128: Security Access Levels

    Password not needed to change settings to view only access. Change Security Access to Full Access: User can switch to full access ● mode. Password is needed to change settings to full access. 3-104 48iQTL Instruction Manual Thermo Scientific...
  • Page 129 Operation Settings Change Full Access Security Password: Full access password can have a ● blank value or user selected password. Thermo Scientific 48iQTL Instruction Manual 3-105...
  • Page 130: Change Security To View Only Access

    Cancel: Exit screen. ● Note To change security access from view only access to full access, a keypad will be displayed where the user can enter full access password. ▲ 3-106 48iQTL Instruction Manual Thermo Scientific...
  • Page 131: Change Full Access Security Password

    The Change Full Access Security Password screen allows the user to set a new password for allowing full access. Security Password Home Screen>Settings>Security Access Levels>Change Full Access Security Password Home Screen>Settings>Security Access Levels>Change Full Access > Security Password Continue Thermo Scientific 48iQTL Instruction Manual 3-107...
  • Page 132 Commit New Security Access Password Change: Commits new security ● password. Cancel and Return to the Home Screen: Exits screen and returns to the ● Home Screen without changing password. 3-108 48iQTL Instruction Manual Thermo Scientific...
  • Page 133: Usb Drive

    Firmware Update Via USB Drive: If USB is mounted, user can update ● instrument firmware. Download Data To USB Drive: If USB is mounted, user can ● download/upload information. Change USB Password: User can change the USB password. ● Thermo Scientific 48iQTL Instruction Manual 3-109...
  • Page 134: Firmware Update Via Usb Drive

    The Firmware Update Via USB Drive screen contains the following information: Update Firmware: User chooses firmware file from USB and updates ● instrument firmware. Instrument reboots when update is finished. Exit: User exits without updating firmware. ● 3-110 48iQTL Instruction Manual Thermo Scientific...
  • Page 135: Download Data To Usb Drive

    USB. Restore: Allows the user to upload the configuration files from the USB ● to the instrument. Download All Data: Downloads all reports, logs, histories, and backup ● information. Thermo Scientific 48iQTL Instruction Manual 3-111...
  • Page 136 If a USB password has been previously set, you will be prompted to enter the USB password to continue. Press Enter to continue. 2. To continue, select the OK button. 3. The USB Drive will display. Select Download Data To USB Drive. 3-112 48iQTL Instruction Manual Thermo Scientific...
  • Page 137 “Success!” message and display the file name as it is stored on the USB flash drive. (The file name format is the instrument serial number, name of download, followed by a date/time stamp.) Remove the USB flash drive and select the OK button to continue. Thermo Scientific 48iQTL Instruction Manual 3-113...
  • Page 138: Change Usb Password

    Settings Change USB Password The Change USB Password screen allows the user to set a new password for accessing USB. Home Screen>Settings>USB Drive>Change USB Password Home Screen>Settings>Security Access Levels>Change Standard Access > Password Continue 3-114 48iQTL Instruction Manual Thermo Scientific...
  • Page 139 ● password for spelling confirmation. Commit New USB Password Change: Commits new USB password. ● Cancel and Return to the Home Screen: Exits screen and returns to the ● Home Screen without changing password. Thermo Scientific 48iQTL Instruction Manual 3-115...
  • Page 140: User Contact Information

    ● Business Address ● Business Shipping Address ● To: User Email Address ● CC: User Email Address 110 ● User Phone Number ● Alternate User Phone Number ● Shelter / Lab Phone Number ● 3-116 48iQTL Instruction Manual Thermo Scientific...
  • Page 141: Update Bootloader

    If button is greyed out, no update is needed. Home Screen>Settings>Update Bootloader The Update Bootloader screen contains the following information: Reboot and Update: Update bootloader and reboot instrument. ● Exit: Exits screen. ● Thermo Scientific 48iQTL Instruction Manual 3-117...
  • Page 143: Chapter 4 Calibration

    Zero Air Generator Since the 48iQTL instrument is equipped with an internal CO scrubber, a separate source of zero air is not required for routine single point calibrations. However, dilution air to generate span gas from a higher concentration cylinder requires zero air that is free of contaminants which will cause a detectable response on the CO analyzer.
  • Page 144: External Flow Meter(S) And Controller(S)

    So cylinder zero air should be scrubbed of the residual CO prior to its use in the 48iQTL as a dilution gas or a zero standard. Room air which has been scrubbed of CO can also be used as the zero air source.
  • Page 145: Calibration

    Calibration Use the following procedure to calibrate the instrument. Connect the Instrument Connect the instrument and the calibration equipment as shown in Figure 4–1. Figure 4–1. Calibration Flow Schematic Thermo Scientific 48iQTL Instruction Manual...
  • Page 146: Detector Gain

    The user can set the Detector Gain to factory defaults by pressing on the cell labeled Default Gain located in the Detector Gain column. Home Screen>Measurement Settings>Advanced Measurement Settings>Optical Bench Settings>Continue To Detector Gain Calibration 48iQTL Instruction Manual Thermo Scientific...
  • Page 147: Initial S/R

    The user can set the Initial S/R to factory defaults by pressing on the cell labeled Default S/R located in the Initial S/R column. Home Screen>Settings>Measurement Settings>Advanced Measurement Settings>Optical Bench Settings>Continue To Initial S/R Calibration Thermo Scientific 48iQTL Instruction Manual...
  • Page 148: Multipoint Calibration

    Cal point coefficients. ▲ To set default values, from the Home Screen select Calibration>Advanced Calibration>Multipoint Calibration. From Multipoint Calibration, select Reset Points to Default Values. The instrument will automatically reset to default values. 48iQTL Instruction Manual Thermo Scientific...
  • Page 149 3. From the Home Screen press Calibration>Advanced Calibration>Multipoint Calibration>Point 1. 4. Enter the span concentration being introduced to the instrument. Press Calibrate. The instrument will perform a series of calculations and save the new parameters. Thermo Scientific 48iQTL Instruction Manual...
  • Page 150: High And Low Multipoint Calibration

    Calibration Calibration Frequency In order to generate data of the highest confidence, it is recommended that a multipoint calibration be performed: every three months ● any time any major disassembly of components is performed ● 48iQTL Instruction Manual Thermo Scientific...
  • Page 151: Calibrate Background

    The Target Concentration button will read 0.000. The Calculated Background button will display the background needed to make the current CO concentration go to zero. 5. Press Calibrate to set the CO reading to zero and to save the new background. Thermo Scientific 48iQTL Instruction Manual...
  • Page 152: Calibrate Span Coefficient

    4. Enter the CO calibration gas concentration using the pushbuttons, and then press Calibrate to calibrate the CO reading to the CO calibration gas. 5. Record the CO concentration and the instrument’s response. 4-10 48iQTL Instruction Manual Thermo Scientific...
  • Page 153 Calibration Calibration Thermo Scientific 48iQTL Instruction Manual 4-11...
  • Page 154: Periodic Zero And Span Checks

    For detailed guidance in setting up a quality assurance program, refer to the Code of Federal Regulations and the EPA Handbook on Quality Assurance. 4-12 48iQTL Instruction Manual Thermo Scientific...
  • Page 155: Manual Calibration

    Adjust Background allows the user to change zero background. The second button called Adjusted Concentration shows what the new CO concentration would be based on the changed zero background. Press the Calibrate button to save the adjusted zero background value. Thermo Scientific 48iQTL Instruction Manual 4-13...
  • Page 156: Adjust Span Coefficient

    Span Coefficient (single range mode) Reset Bkg to 0.000 The Reset Bkg to 0.000 and Span Coef to 1.000 screen allows the user to reset the calibration configuration values to factory defaults. and Span Coef to 1.000 4-14 48iQTL Instruction Manual Thermo Scientific...
  • Page 157: Zero/Span Schedule

    The span duration button is used to set how long the span gas is sampled by the instrument. The purge duration button is used to set how long the purge period will be after doing Thermo Scientific 48iQTL Instruction Manual 4-15...
  • Page 158: Schedule Averaging Time

    Section 12 also provides information on “Calibration of Primary and Secondary Standards for Flow Measurements”. Specific information on certification of concentration standards is given in EPA Traceability Protocol for Assay and Certification of Gaseous Calibration Standards, available at www.epa.gov. 4-16 48iQTL Instruction Manual Thermo Scientific...
  • Page 159: Chapter 5 Maintenance

    2. Flush the filter with warm water and let dry (a clean, oil-free purge will help the drying process) or blow the filter clean with compressed air. 3. Re-install the filter and fan guard. Thermo Scientific 48iQTL Instruction Manual 5-1...
  • Page 160: Pump Rebuilding

    4. Lift the head plate and the intermediate plate off the housing. 5. Hold the pump with one hand, so that the diaphragm is pointing downwards. Lift the diaphragm by the opposing side edges, grasp it and unscrew it in the counter-clockwise direction. 48iQTL Instruction Manual Thermo Scientific...
  • Page 161 12. Place the head plate on the intermediate plate, in the position indicated by the drawing line; gently tighten the four screws, evenly and diagonally (if a torque screwdriver is available: torque about 0.30 Nm). 13. Let the pump run. Thermo Scientific 48iQTL Instruction Manual...
  • Page 162: Figure 5-2. Pump Rebuilding

    Maintenance Pump Rebuilding Figure 5–2. Pump Rebuilding 48iQTL Instruction Manual Thermo Scientific...
  • Page 163: Leak Test

    4. Squeeze trigger until gauge reads to pull in 10 in Hg. 5. Observe vacuum gauge for stable reading for 5 minutes. If reading remains at 10 in Hg, no leak is present. Note Acceptable leak rate is .5 in Hg over 10 minutes. ▲ Thermo Scientific 48iQTL Instruction Manual...
  • Page 164: Cleaning The Optics

    1. Turn off power and disconnect power line. 2. Using 9/64 hex drive, remove the field mirror by removing the four #8- 32 socket cap head screws holding it to the main bench. Figure 5–3. Removing the Field Mirror 48iQTL Instruction Manual Thermo Scientific...
  • Page 165 Dry by blowing clean dry air over the mirror. 4. Reassemble following the above procedure in reverse. It is not necessary to realign any mirror following cleaning. 5. Calibrate the instrument. See the “Calibration” chapter in this manual. Thermo Scientific 48iQTL Instruction Manual...
  • Page 166: Ir Source Replacement

    There is no light output ● After cleaning the optics, the IR light intensities remain below 100,000 ● For more information on replacing the IR Source, see “IR Source Replacement” on page 7-45. 48iQTL Instruction Manual Thermo Scientific...
  • Page 167: Chapter 6 Troubleshooting

    Guide experience an instrument problem. It also lists 48iQTL specific troubleshooting information and alarm messages you may see on the graphics display and provides recommendations about how to resolve the alarm condition.
  • Page 168 Low line voltage Check to see if line voltage is within specifications. Alarm – Instrument Case fan failure Replace fan if not operating properly. Temperature Ambient heater Replace ambient heater assembly if not failure operating properly. 48iQTL Instruction Manual Thermo Scientific...
  • Page 169 Check that DMC cable is connected connection properly. Reseat if needed. Defective Check for other alarms, as it is possible component that another component of that DMC is drawing too much current. Defective DMC Replace DMC board. Thermo Scientific 48iQTL Instruction Manual...
  • Page 170 Replace board as needed. board Alarm – Motor Speed too Cable Check connection from motor to DMC high connection board. Reseat if needed. Defective Heater Replace motor as needed. Defective DMC Replace board as needed. board 48iQTL Instruction Manual Thermo Scientific...
  • Page 171 Alarm – IR Detector Bias Cable Check the cable connection from the IR High connection Source to the DMC board. Defective IR Replace IR Source as needed. Source Defective DMC Replace board as needed. board Thermo Scientific 48iQTL Instruction Manual...
  • Page 172 DMC board. Alarm – Ambient Heater Ambient heater Replace ambient heater assembly if not Status failure operating properly. Cable Check the cable connection from the connection heater assembly to the DMC board. 48iQTL Instruction Manual Thermo Scientific...
  • Page 173: Chapter 7 Servicing

    Chapter 7 Servicing This chapter describes the periodic servicing procedures that should be performed on the instrument to ensure proper operation and explains how to replace the 48iQTL subassemblies. Safety Read the safety precautions before beginning any procedures in this chapter.
  • Page 174: Figure 7-1. Properly Grounded Antistatic Wrist Strap

    When the instrument is unplugged, the chassis is not at earth ground. ▲ Figure 7–1. Properly Grounded Antistatic Wrist Strap Note Ground to unpainted case or outlet as shown. ▲ 48iQTL Instruction Manual Thermo Scientific...
  • Page 175: Firmware Updates

    For more information on installing new firmware, see “Installing New Firmware” in the iQ Series Communications manual. Replacement For a complete list of spare parts, visit the company website at: Parts List www.thermofisher.com/48iQTL Refer to Figure 7–2 Figure 7–3 to identify the component location. Thermo Scientific 48iQTL Instruction Manual...
  • Page 176: Figure 7-2. 48Iqtl Component Layout Top View

    Servicing Replacement Parts List Figure 7–2. 48iQTL Component Layout Top View Figure 7–3. 48iQTL Component Layout Side View 48iQTL Instruction Manual Thermo Scientific...
  • Page 177: Fuse Replacement

    1. Turn instrument OFF and unplug the power cord. 2. Remove fuse drawer, located on the AC power connector. 3. If either fuse is blown, replace both fuses. 4. Insert fuse drawer and reconnect power cord. Figure 7–4. Replacing the Fuses Thermo Scientific 48iQTL Instruction Manual...
  • Page 178: Filter Replacement

    2. Starting with top right corner, pull out to remove fan cover. Figure 7–5. Start with Top Right Corner of Fan Cover 3. Replace filter and snap fan cover back in place. Figure 7–6. Removing the Fan Cover 48iQTL Instruction Manual Thermo Scientific...
  • Page 179: Fan Replacement

    3. Starting with top right corner, pull out to remove fan cover. 4. Unhook the four latches of the fan cover. 5. Unfasten the four 6-32 screws from the fan housing. 6. Replace fan and reassemble in reverse order. Thermo Scientific 48iQTL Instruction Manual...
  • Page 180: Figure 7-8. Replacing The Fan

    Servicing Fan Replacement Figure 7–8. Replacing the Fan 48iQTL Instruction Manual Thermo Scientific...
  • Page 181: Measurement Side Removal

    1. Turn instrument OFF, unplug power cord, and remove the cover (Figure 2–1). 2. Unplug the fan cable J12 on optical bench DMC board. Figure 7–9. Unplugging the Fan Cable 3. Unplug DMC cable (Figure 7–10). Thermo Scientific 48iQTL Instruction Manual...
  • Page 182: Figure 7-10. Unplugging The Dmc Cable

    4. Gripping from the top corners of the front panel and pull outwards. 5. Using #2 Phillips drive, remove three 8-32 flat head screws (Figure 7– 11). Figure 7–11. Unfasten Hardware Front for Measurement Side Removal 7-10 48iQTL Instruction Manual Thermo Scientific...
  • Page 183: Figure 7-12. Unfasten Hardware Rear For Measurement Side Removal

    7. Unfasten captive hardware. 8. Using #2 Phillips drive, remove two 8-32 flat head screws. 9. Pull measurement side out. 10. Replace in reverse order. Figure 7–12. Unfasten Hardware Rear for Measurement Side Removal Thermo Scientific 48iQTL Instruction Manual 7-11...
  • Page 184: Lcd Module Replacement

    2. Gripping from the top corners of the front panel and pull outwards. 3. Using 1/4 wrench, unfasten four #4-40 nuts (Figure 7–13). Figure 7–13. Replacing the LCD Module 4. Remove cover. 5. Unplug LCD cables from backside of board. 6. Pull board off the standoffs. 7-12 48iQTL Instruction Manual Thermo Scientific...
  • Page 185: Figure 7-14. Remove Electrical Cables From Lcd

    Servicing LCD Module Replacement Figure 7–14. Remove Electrical Cables from LCD 7. Replace LCD module and reassemble in reverse order. Thermo Scientific 48iQTL Instruction Manual 7-13...
  • Page 186: I/O Replacement

    Use the following procedure to replace the I/O boards. 1. Turn instrument OFF, unplug power cord, and remove the cover (Figure 2–1). 2. Swing arm open. Figure 7–15. I/O Replacement, Arm 3. Pull board upwards. Figure 7–16. I/O Replacement, Remove Board 7-14 48iQTL Instruction Manual Thermo Scientific...
  • Page 187: Figure 7-17. I/O Replacement, Install

    4. During install, make sure to align cutout circular to keyway. 5. Insert board downwards. Figure 7–17. I/O Replacement, Install 6. Close arm. Make sure expansion bracket aligns to the inside of the rectangular cutouts. Figure 7–18. I/O Replacement, Close Arm Alignment Thermo Scientific 48iQTL Instruction Manual 7-15...
  • Page 188: Peripherals Support Board And System Controller Board Replacement

    1. Turn instrument OFF, unplug power cord, and remove the cover Controller Board (Figure 2–1). Replacement 2. Pull tab out (two per board). 3. Pull board out. Figure 7–19. Replacing the Peripherals Support Board 4. Replace board and reassemble in reverse order. 7-16 48iQTL Instruction Manual Thermo Scientific...
  • Page 189: Dmc Pressure And Flow Board

    (Figure 2–1). 2. Unplug cables from the pressure and flow board. Figure 7–20. Flow Pressure Board, Disconnect DMC Cable 3. Disconnect plumbing. 4. Using 7/16 hex drive, unfasten four #6-32 socket cap head screws. Thermo Scientific 48iQTL Instruction Manual 7-17...
  • Page 190: Figure 7-21. Flow Pressure Board, Screws

    Servicing DMC Pressure and Flow Board Figure 7–21. Flow Pressure Board, Screws 5. Replace board and reassemble in reverse order. 7-18 48iQTL Instruction Manual Thermo Scientific...
  • Page 191: Pump Replacement

    3. Twist opposite direction to unlock tube clamps. Note Push in tube clamp to lock. ▲ 4. Disconnect tubing from pump. 5. Using #2 Phillips drive, unfasten two captive hardware. 6. Slide pump left until keyway meets opening. Thermo Scientific 48iQTL Instruction Manual 7-19...
  • Page 192: Figure 7-22. Remove Pump, Disconnect And Unfasten

    Servicing Pump Replacement Figure 7–22. Remove Pump, Disconnect and Unfasten 7. Pull pump outwards. Note When installing pump, make sure the pump keyway opening goes over the keyway. ▲ 7-20 48iQTL Instruction Manual Thermo Scientific...
  • Page 193: Figure 7-23. Pump Removal, Keyway

    Servicing Pump Replacement Figure 7–23. Pump Removal, Keyway 8. Using #1 Phillips drive, remove two M3 screws. Figure 7–24. Pump replacing, Unfasten Screws 9. Replace pump and reassemble in reverse order. Thermo Scientific 48iQTL Instruction Manual 7-21...
  • Page 194: Capillary Cleaning And/Or Replacement

    1. Turn instrument OFF, unplug power cord, and remove the cover (Figure 2–1). 2. Disconnect the plumbing. 3. Using #2 Phillips drive, unfasten captive hardware. Figure 7–25. Remove Capillary, Disconnect and Unfasten 4. Slide capillary plate upwards clearing the partition panel keyway. 7-22 48iQTL Instruction Manual Thermo Scientific...
  • Page 195: Figure 7-26. Capillary Plate, Keyway

    Servicing Capillary Cleaning and/or Replacement Figure 7–26. Capillary Plate, Keyway 5. Using 7/64 hex drive, remove four #6-32 socket cap head screws. 6. Pull apart the capillary blocks. Figure 7–27. Capillary and O-Ring Replace Thermo Scientific 48iQTL Instruction Manual 7-23...
  • Page 196: Capillary O-Ring Replacement

    1. Turn instrument OFF, unplug power cord, and remove the cover (Figure 2–1). 2. Unplug all electrical shown J9, J10, J24, J25, J26, and ground. 3. Unfasten captive hardware. 4. Slide power supply left, clearing three case floor plate keyways. 7-24 48iQTL Instruction Manual Thermo Scientific...
  • Page 197: Figure 7-28. Removing Power Supply

    Servicing Power Supply Replacement Figure 7–28. Removing Power Supply 5. Pull power supply up. 6. Replace power supply and reassemble in reverse order. Figure 7–29. Replacing Power Supply Thermo Scientific 48iQTL Instruction Manual 7-25...
  • Page 198: Step Pol Board Replacement

    1. Turn instrument OFF, unplug power cord, and remove the cover (Figure 2–1). 2. Unplug Step POL power cable J4. 3. Unplug Step POL signal cable J2. 4. Unplug pump cable J7. 5. Unfasten captive hardware. Figure 7–30. Unplug and Unfasten Step POL Board 7-26 48iQTL Instruction Manual Thermo Scientific...
  • Page 199: Figure 7-31. Clear Partition Keyway Step Pol Board

    7–32). If replacing optional step board 2, make sure switch 1 is pointed towards ON and switch 2 is pointed away from ON (Figure 7–33). 8. Replace Step POL board and reassemble in reverse order. Thermo Scientific 48iQTL Instruction Manual 7-27...
  • Page 200: Figure 7-32. Step Pol Board 1 Switch Settings

    Servicing Step POL Board Replacement Figure 7–32. Step POL Board 1 Switch Settings Figure 7–33. Optional Step POL Board 2 Switch Settings 7-28 48iQTL Instruction Manual Thermo Scientific...
  • Page 201: Dmc Optical Bench

    Optical Bench removal ● Optical switch replacement, purge wheel housing removal, filter wheel ● replacement, motor replacement, filter wheel and motor alignment Heater Assembly removal ● Pre-Amp/Detector Assembly removal ● IR Source removal ● Thermo Scientific 48iQTL Instruction Manual 7-29...
  • Page 202: Optical Bench Removal

    2. Unplug DMC cable J3 (top). Unplug heater cable J7. Unplug case fan cable J12. 3. Disconnect plumbing as shown. 4. Using a #2 Phillips drive, unfasten six captive hardware. Figure 7–35. DMC Removal from Instrument 7-30 48iQTL Instruction Manual Thermo Scientific...
  • Page 203: Optical Switch Replacement

    Use the following procedure to replace the optical switch. Replacement Equipment required: Hex drive, 5/32 Phillips drive, #1 1. Unplug J8 (motor) and J4 (opt). 2. Unhook motor and opt connections. Figure 7–36. Motor and Filter Removal pt 1 Thermo Scientific 48iQTL Instruction Manual 7-31...
  • Page 204: Figure 7-37. Motor And Filter Removal (Pt 2)

    Servicing DMC Optical Bench 3. Unplug electrical connections. 4. Using a 5/32 hex drive, remove three #10-32 socket cap head screws and lockwashers. Figure 7–37. Motor and Filter Removal (pt 2) 7-32 48iQTL Instruction Manual Thermo Scientific...
  • Page 205: Figure 7-38. Motor And Filter Optical Switch Replacement (Pt 3)

    5. Using #1 Phillips drive, unfasten two #4-40 flat head screws to remove the optical cover. 6. Using #1 Phillips drive, unfasten one #2-56 pan head screw to remove the optical switch. Figure 7–38. Motor and Filter Optical Switch Replacement (pt 3) Thermo Scientific 48iQTL Instruction Manual 7-33...
  • Page 206: Purge Wheel Housing Removal

    Housing Removal Equipment required: Hex drive, 5/32 1. Using 5/32 hex drive, unfasten two #10-32 socket cap head screws. 2. Carefully pull wheel housing apart. Figure 7–39. Purge Housing Removal from Motor Assembly (pt 4) 7-34 48iQTL Instruction Manual Thermo Scientific...
  • Page 207: Filter Wheel Replacement

    5. Using 5/64 hex drive, loosen #8-32 set screw two or three turns only. 6. Carefully slide chopper wheel off motor shaft. 7. Replace filter wheel and align motor and filter wheel. Refer to “Motor and Filter Wheel Alignment” on page 7-39. Thermo Scientific 48iQTL Instruction Manual 7-35...
  • Page 208: Figure 7-40. Filter Wheel

    Servicing DMC Optical Bench Figure 7–40. Filter Wheel 7-36 48iQTL Instruction Manual Thermo Scientific...
  • Page 209: Motor Replacement

    Removal” on page 7-34. 3. Remove filter wheel. Refer to “Filter Wheel Replacement” on page 4. Using 5/64 hex drive, unfasten two #10-32 socket cap head screws. Figure 7–41. Motor Removal from Motor Mount pt 1 Thermo Scientific 48iQTL Instruction Manual 7-37...
  • Page 210: Figure 7-42. Motor Removal

    6. Using 5/64 hex drive, loosen #8-32 set screw two or three turns only. 7. Carefully slide motor out. 8. Replace motor and align filter wheel and motor. Refer to “Motor and Filter Wheel Alignment” on page 7-39. Figure 7–42. Motor Removal 7-38 48iQTL Instruction Manual Thermo Scientific...
  • Page 211: Motor And Filter Wheel Alignment

    4. Make sure 10-32 screws are fastened. If motor is replaced, refer to Figure 7–41. 5. After the filter wheel is installed, spin the wheel and observe that it runs true on the motor shaft. Thermo Scientific 48iQTL Instruction Manual 7-39...
  • Page 212: Figure 7-43. Motor Installation Shaft Alignment

    6. Let instrument sample zero air for about 90 minutes. 7. From the Home screen, choose Settings>Measurement Settings>Advanced Measurement Settings>Optical Bench Settings>Continue to Initial S/R Calibration. 8. Follow steps in Chapter 4, “Calibration” for “Initial S/R”. 7-40 48iQTL Instruction Manual Thermo Scientific...
  • Page 213 1. Unplug electrical connections, J5 heater, J1 preamp, bottom connection J8 motor, and top connection J4 opt. 2. Using a 5/32 hex drive, unfasten four #10-32 socket cap head screws and lockwashers. Figure 7–44. Bench Removal pt 1 Thermo Scientific 48iQTL Instruction Manual 7-41...
  • Page 214: Figure 7-45. Heater Removal

    4. Using #1 Phillips drive, unfasten five #4-40 pan head screws. Figure 7–45. Heater Removal 5. When replacing with new heatsink or thermistor, heatsink compound is required. Apply a thin and even coat of heatsink compound underneath. 7-42 48iQTL Instruction Manual Thermo Scientific...
  • Page 215: Pre-Amp/Detector Assembly Removal

    2. Unplug preamp and source IR cable assembly. Figure 7–46. Preamp Detector Cover Removal pt 1 3. Using a #1 Phillips drive, unfasten two #4-40 pan head screws. 4. Lift preamp cover up. Thermo Scientific 48iQTL Instruction Manual 7-43...
  • Page 216: Figure 7-47. Preamp/Detector Removal

    6. Pull preamp board assembly outwards. Figure 7–47. Preamp/Detector Removal 7. Replace preamp/detector as needed and assemble in reverse order. Note When reinstalling preamp board assembly, make sure the o-ring is seated properly. ▲ 7-44 48iQTL Instruction Manual Thermo Scientific...
  • Page 217: Ir Source Replacement

    Use the following procedure to replace the IR source from DMC. 1. Turn the instrument OFF, unplug the power cord, and remove the cover. 2. Disconnect IR source cable as shown. Figure 7–48. IR Source Cable Disconnect Thermo Scientific 48iQTL Instruction Manual 7-45...
  • Page 218: Figure 7-49. Ir Source Assembly Removal

    Servicing DMC Optical Bench 3. Using a #1 Phillips drive, unfasten two #4-40 pan head screws. 4. Pull IR source assembly upwards. Figure 7–49. IR Source Assembly Removal 5. Reinstall in reverse order. 7-46 48iQTL Instruction Manual Thermo Scientific...
  • Page 219: Heater Assembly

    2. Unplug electrical, unplug heater cable J7. 3. Using a #2 Phillips drive, unfasten captive hardware. Figure 7–50. Heater Assembly Removal pt 1 4. Slide heater assembly left. Note Make sure heater plate clears the edge and stud. ▲ Thermo Scientific 48iQTL Instruction Manual 7-47...
  • Page 220 Figure 7–51. Heater Assembly Removal pt 2 5. Slide heater assembly upwards. Note Make sure heater plate clears the edge. ▲ Figure 7–52. Heater Assembly Removal pt 3 6. Pull away from the partition panel. 7-48 48iQTL Instruction Manual Thermo Scientific...
  • Page 221 Servicing Heater Assembly Figure 7–53. Heater Assembly Removal pt 4 Thermo Scientific 48iQTL Instruction Manual 7-49...
  • Page 222: Heater Assembly Install

    2. Slide heater assembly right. Note Make sure heater plate fits under the stud. ▲ Note Make sure heater plate slides into side pocket. ▲ 3. Using a #2 Phillips drive, fasten captive hardware. 7-50 48iQTL Instruction Manual Thermo Scientific...
  • Page 223 Servicing Heater Assembly Figure 7–55 Heater Assembly Install pt 2 Thermo Scientific 48iQTL Instruction Manual 7-51...
  • Page 224: Sample/Zero Valve Replacement

    2–1). 2. Unplug sample/zero valve (J8) from the Step POL board 1. Figure 7–56. Replacing the Sample/Zero Valve pt 1 3. Disconnect all plumbing. 4. Using a #2 Phillips drive, unfasten two captive hardware. 7-52 48iQTL Instruction Manual Thermo Scientific...
  • Page 225 5. Using a 9/64 hex drive, unfasten two socket cap head screws and remove valve from bracket. 6. Use vice and wrench and to replace fittings. Figure 7–58. Replacing the Sample/Zero Valve pt 3 Thermo Scientific 48iQTL Instruction Manual 7-53...
  • Page 226 Wrap around approximate two times around depending on the thickness of the PTFE tape. Make sure PTFE tape does not overhang the first thread (the lead in thread). ▲ 7-54 48iQTL Instruction Manual Thermo Scientific...
  • Page 227: Optional Manifold Replacement

    2. Unplug three electrical connections (J5, J6, and J8) from the Step POL board 2. Figure 7–59. Replacing the Manifold pt 1 3. Unfasten three nuts. Remove the nuts, front and back ferrules as shown from span, zero in, sample back panel (Figure 7–60). Thermo Scientific 48iQTL Instruction Manual 7-55...
  • Page 228: Figure 7-60. Replacing The Manifold Pt 2

    Figure 7–60. Replacing the Manifold pt 2 4. Disconnect tubing. 5. Using a 9/64 hex drive, unfasten four #8-32 socket cap head screws. Figure 7–61. Replacing the Manifold pt 3 6. Replace the manifold and assemble in reverse order. 7-56 48iQTL Instruction Manual Thermo Scientific...
  • Page 229: Zero Air Scrubber Replacement

    1. Turn the instrument OFF, unplug the power cord, and remove the cover (Figure 2–1). 2. Disconnect plumbing. 3. Unfasten four #8-32 socket cap head screws. Figure 7–62. Zero Air Scrubber Replacement 4. Replace zero air scrubber and reinstall in reverse order. Thermo Scientific 48iQTL Instruction Manual 7-57...
  • Page 231: Chapter 8 System Description

    Chapter 8 System Description The 48iQTL deploys a set of modular subsystems that comprise the total instrument function. The core measurements for concentration are contained in Distributed Measurement and Control (DMC) modules. This chapter describes the function and location of the system components in the module framework, including firmware, electronics, and I/O function.
  • Page 232: Optical Bench Dmc

    Figure 8–1 shows how the filter wheel purge is configured within the instrument for an external purge gas supply. 48iQTL Instruction Manual Thermo Scientific...
  • Page 233: Detector Preamp

    System Description Optical Bench DMC Figure 8–1. 48iQTL Flow Schematic with Filter Wheel Purge Calibration Note The stability of the purge gas is very important. Since this gas is in the same optical path as the sample, changes in the composition of the purge gas can change the calibration of the instrument.
  • Page 234: Optical Bench Dmc Board

    The rear panel fan, which is part of the common electronics in other iQ analyzers is repurposed in the 48iQTL to provide tight control over air flow into the instrument. The rear panel fan is plugged into the optical bench DMC board which controls the fan speed to help maintain a stable instrument temperature.
  • Page 235: Figure 8-2. Common System Interconnect Diagram

    System Description Common Electronics Figure 8–2. Common System Interconnect Diagram Thermo Scientific 48iQTL Instruction Manual...
  • Page 236: Figure 8-3. 48Iqtl System Interconnect Diagram

    System Description Common Electronics Figure 8–3. 48iQTL System Interconnect Diagram 48iQTL Instruction Manual Thermo Scientific...
  • Page 237: Power Supply

    All electronics operate from a universal VDC supply, which is capable of auto-sensing the input voltage and working over all specified operating ranges. The 48iQTL contains a 24 VDC channel for most electronics operation, including the pump and fan, and a 48 VDC channel dedicated specifically for optical bench heating.
  • Page 238: Peripherals Support System

    The pressure differential determines the flow through the capillary. The upstream pressure is the measurement bench pressure, while the downstream pressure is at the pump inlet pressure. 48iQTL Instruction Manual Thermo Scientific...
  • Page 239: Firmware

    SNMP registers, some of which is interfaced with the front panel Graphical User Interface (GUI). In addition to the operating registers, the 48iQTL stores a historical data log in a MySQL database. The memory is provided on the same micro SD card where the operating software resides, for which there is capability to store up to a year of data at 1 minute intervals.
  • Page 241: Chapter 9 Optional Equipment

    Chapter 9 Optional Equipment The 48iQTL is available with the following options: Connecting Several components are available for connecting external devices. External Devices These connection options consist of three plug-in boards: Communication Board ● Analog I/O Board ● Digital I/O Board ●...
  • Page 242: Communication Board

    RS-232 or RS-485 communication. Figure 9–2. RS-232/RS-485 Port Table 9–1. RS-232/RS-485 Port Terminal Assignment Terminal Number Signal Name No Connect RX/RS485_RX_P TX/RS485_TX_N No Connect No Connect RTS/RS485_TX_P CTS/RS485_RX_N No Connect 48iQTL Instruction Manual Thermo Scientific...
  • Page 243: Figure 9-3. Rs-485 External Accessory Port

    Table 9–2. RS-485 External Accessory Port Terminal Assignment Terminal Number Signal Name EXT_RS485_RX_N EXT_RS485_RX_P +5V (Fused @0.4A) +5V (Fused @0.4A) +5V (Fused @0.4A) EXT_RS485_TX_N EXT_RS485_TX_P +24V (Fused @0.4A) +24V (Fused @0.4A) +24V (Fused @0.4A) +24V (Fused @0.4A) +24V (Fused @0.4A) Thermo Scientific 48iQTL Instruction Manual...
  • Page 244: Analog I/O Board

    Figure 9–4. Analog Voltage Inputs Table 9–3. Analog Voltage Inputs Assignment Terminal Number Signal Name Analog In 1 Analog GND Analog In 2 Analog GND Analog In 3 Analog GND Analog In 4 Analog GND 48iQTL Instruction Manual Thermo Scientific...
  • Page 245: Analog Voltage Outputs

    Current Output is 1000 Ω. All Voltage Outputs and Current Outputs are continuously monitored separately for accuracy. The Analog Outputs may be used to control and report parameters pertinent to the analyzers’ measured functions. Figure 9–5. Analog Voltage and Current Thermo Scientific 48iQTL Instruction Manual...
  • Page 246: Analog Output Calibration

    Note When calibrating the current output when using the 0-20 mA scale, the low level will be set to 4 mA due to the inability to adjust the actual current output to below zero. ▲ Calibrate the output channel full scale. ● 48iQTL Instruction Manual Thermo Scientific...
  • Page 247 2. Depending on the output type being used, select either Analog Out Zero Calibration (Voltage) or Analog Out 4.000 mA Calibration (Current). 3. A confirmation screen is presented. Select Continue to proceed with the calibration or Return to Previous Screen. Thermo Scientific 48iQTL Instruction Manual...
  • Page 248 6. After making changes to the output levels, the commit button will turn green. To accept the changes, press the Commit button. To revert to the previous values, press the back button to return to the previous analog output calibration screen. 48iQTL Instruction Manual Thermo Scientific...
  • Page 249 2. Depending on the output type being used, select either Analog Out Full Scale Calibration (Voltage) or Analog Out 20.000 mA Calibration (Current). 3. A confirmation screen is presented. Select Continue to proceed with the calibration or Return to Previous Screen. Thermo Scientific 48iQTL Instruction Manual...
  • Page 250 6. After making changes to the output levels, the commit button will turn green. To accept the changes, press the Commit button. To revert to the previous values, press the back button to return to the previous analog output calibration screen. 9-10 48iQTL Instruction Manual Thermo Scientific...
  • Page 251: Digital I/O Board

    The active state can be user defined in firmware. Logic Low Threshold: 0.8 V ● Logic High Threshold: 2.0 V ● Absolute allowable input voltages: -0.5 to 5.5 V ● Figure 9–6. Digital Inputs Thermo Scientific 48iQTL Instruction Manual 9-11...
  • Page 252: Table 9-5. Digital Inputs Terminal Assignment

    Digital In 6 Digital In 7 Digital In 8 Digital In 9 Digital In 10 Digital In 11 Digital In 12 Digital In 13 Digital In 14 Digital In 15 Digital In 16 COMMON 9-12 48iQTL Instruction Manual Thermo Scientific...
  • Page 253: Digital Relay Switches

    Optional Equipment Connecting External Devices Digital Relay Switches Table 9–6 lists the digital relay switches. Maximum Voltage: 300 VDC ● Maximum Current: 500 mA ● Fuse: 800 mA ● Figure 9–7. Digital Relay Switches Thermo Scientific 48iQTL Instruction Manual 9-13...
  • Page 254: Table 9-6. Digital Relay Switch Assignment

    Relay 2B Relay 3A Relay 3B Relay 4A Relay 4B Relay 5A Relay 5B Relay 6A Relay 6B Relay 7A Relay 7B Relay 8A Relay 8B Relay 9A Relay 9B Relay 10A Relay 10B 9-14 48iQTL Instruction Manual Thermo Scientific...
  • Page 255: Valve Driver Outputs

    Actual Output Voltage: 2224 VDC ● Maximum Current: 300 mA ● Both positive and negative outputs are protected from over voltage and ● over current by 500 mA fuses. Figure 9–8. Valve Driver Outputs Thermo Scientific 48iQTL Instruction Manual 9-15...
  • Page 256: Table 9-7. Valve Driver Outputs Assignment

    Valve Drive 7+ Valve Drive 7- Valve Drive 8+ Valve Drive 8- Note Intended for 24 V valves. These outputs will also drive any DC load of 2224 VDC, up to 300 mA. ▲ 9-16 48iQTL Instruction Manual Thermo Scientific...
  • Page 257: Internal Zero/Span Assembly

    It may be necessary to use an atmospheric dump bypass plumbing arrangement to accomplish this. Figure 9–9 shows how this option is integrated with the instrument. Figure 9–9. 48iQTL Flow Schematic with Zero Span Thermo Scientific 48iQTL Instruction Manual 9-17...
  • Page 258: Ptfe Particulate Filter

    PTFE Particulate Filter PTFE Particulate A 510 micron pore size, two-inch diameter PTFE element is available for the 48iQTL. This filter should be installed just prior to the SAMPLE Filter bulkhead. When using a filter, all calibrations and span checks must be performed through the filter.
  • Page 259: Appendix A Safety, Warranty, And Weee

    ▲ qualified service personnel only.   Equipment Damage Do not attempt to lift the analyzer by the cover or other ▲ external fittings. This adjustment should only be performed by an instrument ▲ service technician. Thermo Scientific 48iQTL Instruction Manual...
  • Page 260: Warranty

    Products for which Buyer has requested warranty services are not covered by the warranty hereunder, Buyer shall pay or reimburse Seller for all costs of investigating and responding to such request at Seller's then 48iQTL Instruction Manual Thermo Scientific...
  • Page 261 OR WRITTEN, WITH RESPECT TO THE PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. SELLER DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR-FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT. Thermo Scientific 48iQTL Instruction Manual...
  • Page 262: Weee Compliance

    Description Marking of electrical and electronic equipment which applies to electrical and electronic equipment falling under the Directive 2002/96/EC (WEEE) and the ▲ equipment that has been put on the market after 13 August 2005. 48iQTL Instruction Manual Thermo Scientific...
  • Page 263 Figure 1–1. 48iQTL Front ..................1-2   Figure 1–2. 48iQTL Flow Schematic ..............1-4   Figure 1–3. 48iQTL Flow Schematic with Zero Span Manifold ......1-4   Figure 1–4. Bench Mount Assembly (dimensions in inches [mm]) ....1-6   Figure 1–5. Rack Mount Assembly (dimensions in inches [mm]) ...... 1-7  ...
  • Page 264 Figure 7–51. Heater Assembly Removal pt 2 ........... 7-48   Figure 7–52. Heater Assembly Removal pt 3 ........... 7-48   Figure 7–53. Heater Assembly Removal pt 4 ........... 7-49   Figure 7–54. Heater Assembly Install pt 1 ............7-50   48iQTL Instruction Manual Thermo Scientific...
  • Page 265: List Of Tables

    Figure 7–61. Replacing the Manifold pt 3 ............7-56   Figure 7–62. Zero Air Scrubber Replacement ..........7-57   Figure 8–1. 48iQTL Flow Schematic with Filter Wheel Purge ......8-3   Figure 8–2. Common System Interconnect Diagram ......... 8-5  ...
  • Page 267: Gnu Lesser General Public License

    Our General Public Licenses are designed to make sure that you have the freedom to distribute copies of free software (and charge for this service if you wish); that you receive source code or can get it if you want it; Thermo Scientific 48iQTL Instruction Manual...
  • Page 268 We call this license the "Lesser" General Public License because it does Less to protect the user's freedom than the ordinary General Public License. It also provides other free software developers Less of an advantage over competing non-free programs. These disadvantages are the reason we use 48iQTL Instruction Manual Thermo Scientific...
  • Page 269 Library or any derivative work under copyright law: that is to say, a work containing the Library or a portion of it, either verbatim or with modifications and/or translated straightforwardly into another language. (Hereinafter, translation is included without limitation in the term "modification".) Thermo Scientific 48iQTL Instruction Manual...
  • Page 270 Therefore, Subsection 2d requires that any application-supplied function or table used by this function must be optional: if the application does not supply it, the square root function must still compute square roots.) 48iQTL Instruction Manual Thermo Scientific...
  • Page 271 5. A program that contains no derivative of any portion of the Library, but is designed to work with the Library by being compiled or linked with it, is Thermo Scientific 48iQTL Instruction Manual...
  • Page 272 Library, with the complete machine-readable "work that uses the Library", as object code and/or source code, so that the user can modify the Library and then relink to produce a modified executable containing the modified Library. (It is 48iQTL Instruction Manual Thermo Scientific...
  • Page 273 Give prominent notice with the combined library of the fact that part of it is a work based on the Library, and explaining where to find the accompanying uncombined form of the same work. Thermo Scientific 48iQTL Instruction Manual...
  • Page 274 48iQTL Instruction Manual Thermo Scientific...
  • Page 275 SHOULD THE LIBRARY PROVE DEFECTIVE, YOU ASSUME THE COST OF ALL NECESSARY SERVICING, REPAIR OR CORRECTION. 16. IN NO EVENT UNLESS REQUIRED BY APPLICABLE LAW OR AGREED TO IN WRITING WILL ANY COPYRIGHT HOLDER, OR Thermo Scientific 48iQTL Instruction Manual...
  • Page 276 You should also get your employer (if you work as a programmer) or your school, if any, to sign a "copyright disclaimer" for the library, if necessary. Here is a sample; alter the names: C-10 48iQTL Instruction Manual Thermo Scientific...
  • Page 277 (iii) accessors; and (iv) small macros, templates and inline functions of five lines or less in length. Furthermore, you are not required to apply this additional permission to a modified version of the Library. Thermo Scientific 48iQTL Instruction Manual C-11...
  • Page 280 Ph: +91 22 4157 8800 Ph: +86 10 84193588 Fax: +44 1606 548711 orders.aqi@thermofisher.com india@thermofisher.com info.eid.china@thermofisher.com sales.epm.uk@thermofisher.com Find out more at thermofisher.com/48iQTL © 2018 Thermo Fisher Scientific Inc. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. EPM_119014-00...