Thermo Scientific 49i Instruction Manual page 78

Uv photometric o3 analyzer
Table of Contents

Advertisement

Operation
Instrument Controls Menu
Instrument ID
Communication Protocol
3-42
Model 49i Instruction Manual
Press
to save the new baud rate.
BAUD RATE:
CURRENTLY:
SET TO:
RANGE
AVG
The Instrument ID screen allows the operator to edit the instrument ID.
The ID is used to identify the instrument when using the C-Link or
MODBUS protocols to control the instrument or collect data. It may be
necessary to edit the ID number if two or more of the same instrument are
connected to one computer. Valid Instrument ID numbers are from 0 to
127. The Model 49i has a default Instrument ID of 49. For more
information about the Instrument ID, see Appendix B "C-Link Protocol
Commands" or Appendix C "MODBUS Protocol".
In the Main Menu, choose Instrument Controls > Communication
Settings > Instrument ID.
Use
and
Press
to save the new instrument ID.
INSTRUMENT ID:
CURRENTLY:
SET TO:
RANGE
AVG
The Communication Protocol screen is used to change the instrument
communication protocol for serial communications.
In the Main Menu, choose Instrument Controls > Communication Settings
> Communication Protocol.
9600
19200
?
CHANGE VALUE
SAVE VALUE
DIAGS ALARM
to increment or decrement the ID value.
49
50 ?
CHANGE VALUE
SAVE VALUE
DIAGS ALARM