Motherboard Replacement - Thermo Scientific 49i Instruction Manual

Uv photometric o3 analyzer
Table of Contents

Advertisement

Servicing

Motherboard Replacement

Motherboard
Replacement
7-20
Model 49i Instruction Manual
4. Using the nut driver, remove the two standoffs securing the board to the
rear panel
(Figure
7-8).
5. Pop off the digital output board from the mounting studs and remove
the board.
6. To install the digital output board, follow previous steps in reverse.
Use the following procedure to replace the motherboard
Equipment Required:
Motherboard
Philips screwdriver
Nut driver, 3/16-inch
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the I/O expansion board (optional), if used. See the "I/O
Expansion Board Replacement" procedure in this chapter.
3. Remove the digital output board. See the "Digital Output Board
Replacement" procedure in this chapter.
4. Unplug all connectors from the motherboard. Note connector locations
to facilitate reconnection.
5. Using the nut driver, remove the eight standoffs securing the board to
the rear panel.
(Figure
7-7).
Thermo Fisher Scientific