C-Link Protocol Commands

Diagnostics

Diagnostics
B-16
Model 49i Instruction Manual
Send:
set alarm trig conc no 1
Receive:
set alarm trig conc no 1 ok
Table B-3. Alarm Trigger Values
Value
Alarm Trigger
0
Floor
1
Ceiling
cell a int
cell b int
The example that follows reports that the lamp intensity in cell A is 98,425
Hz.
Send:
cell a int
Receive:
cell a int 98425 Hz
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt mb
Receive:
diag volt mb 24.1 14.9 4.9 3.2 -3.2
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, positive 3.3, and positive 15. Each
voltage value is separated by a space.
Send:
diag volt mib
Receive:
diag volt mib 24.1 14.9 -14.9 4.9 3.2 14.9
diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt iob
Receive:
diag volt iob 24.1 4.9 3.2 -3.2